Reproducibility of Coronary Optical Coherence Tomography for Lumen and Length Measurements in Humans (The CLI-VAR [Centro per la Lotta contro l'Infarto-VARiability] Study) - 26/09/12

Doi : 10.1016/j.amjcard.2012.05.047 
Silvio Fedele, MD a, Giuseppe Biondi-Zoccai, MD a, Piotr Kwiatkowski, MD a, Luca Di Vito, MD a, Michele Occhipinti, MD a, Alberto Cremonesi, MD b, Mario Albertucci, MD a, Laura Materia, PharmD a, Giulia Paoletti, RT a, Francesco Prati, MD a, c,
a Centro per la Lotta contro l'Infarto Foundation, Rome, Italy 
b Sansavini Foundation, Cotignola, Italy 
c Division of Cardiology, San Giovanni Hospital, Rome, Italy 

Corresponding author: Tel: (+39) 0677208915; fax: (+39) 0645443331

Riassunto

Frequency-domain optical coherence tomography (FD-OCT) is becoming a useful diagnostic tool for coronary imaging for quantitative coronary analysis. Second-generation FD-OCT produces detailed coronary lumen images. However, the reproducibility of coronary measurements using FD-OCT in humans has not been thoroughly explored. Our goal was to determine the intraobserver, interobserver, and interpullback reproducibility of the in vivo FD-OCT measurements of the lumen area and/or lesion length. Twenty-five patients undergoing coronary angioplasty were included. In all subjects, FD-OCT pullbacks (20 mm/s) were acquired twice from the same coronary segment different from the target lesion, at an interval of 5 minutes, with no other intervention. A total of 9,396 cross-sectional lumen area frames and the relative coronary lesion length of each pullback were analyzed off-line with dedicated software by 2 independent expert readers (A and B). We compared the lumen area and length measurements as follows: pullback 1, read by reader A twice at an interval of 7 days (intraobserver analysis); pullback 1, independently read by readers A and B (interobserver comparison); and pullback 1 versus pullback 2, read by reader A (interpullback comparison). The per-segment and per-frame analyses showed very high and significant correlation coefficients for the interobserver, intraobserver, and interpullback comparisons for the lumen area and lesion length (R ≥0.95 and p <0.001 in all cases). Accordingly, the Bland-Altman estimates of bias showed nonsignificant differences in the interobserver, intraobserver, and interpullback comparisons at all levels, with average biases never >0.150 mm2 for the lumen area or 0.200 mm for the lesion length. In conclusion, coronary imaging using FD-OCT showed excellent reproducibility, with low intraobserver, interobserver, and interpullback variability for both lumen area and lesion length measurements in humans. Thus, FD-OCT can be proposed for precise analysis in the catheterization laboratory to guide decision making and in clinical trials focusing on imaging end points.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2012  Elsevier Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 110 - N° 8

P. 1106-1112 - ottobre 2012 Ritorno al numero
Articolo precedente Articolo precedente
  • Usefulness of Epicardial Adipose Tissue as Predictor of Cardiovascular Events in Patients With Coronary Artery Disease
  • Felipe N. Albuquerque, Virend K. Somers, Gustavo Blume, William Miranda, Yoel Korenfeld, Andrew D. Calvin, Rui Qin, Francisco Lopez-Jimenez
| Articolo seguente Articolo seguente
  • Comparison of Bare-Metal Stents and Drug-Eluting Stents in Coronary Ostial Lesions (from the National Heart, Lung, and Blood Institute Dynamic Registry)
  • Samip Vasaiwala, Helen Vlachos, Faith Selzer, Oscar Marroquin, Suresh Mulukutla, J. Dawn Abbott, David O. Williams

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.