Abbonarsi

Feasibility and accuracy of a comprehensive multidetector computed tomography acquisition for patients referred for balloon-expandable transcatheter aortic valve implantation - 06/08/11

Doi : 10.1016/j.ahj.2011.03.003 
Gianluca Pontone, MD a, , Daniele Andreini, MD a, Antonio L. Bartorelli, MD, FACC a, b, Andrea Annoni, MD a, Saima Mushtaq, MD a, Erika Bertella, MD a, Alberto Formenti, MD a, Sarah Cortinovis a, Francesco Alamanni, MD a, b, Melissa Fusari, MD a, Veronica Bona a, Gloria Tamborini, MD a, Manuela Muratori, MD a, Giovanni Ballerini, MD a, Cesare Fiorentini, MD a, b, Paolo Biglioli, MD a, Mauro Pepi, MD a
a Centro Cardiologico Monzino, Istituto di Ricovero e Cura a Carattere Scientifico, Milan, Italy 
b Department of Cardiovascular Sciences, University of Milan, Milan, Italy 

Reprint requests: Gianluca Pontone, MD, Via C. Parea 4, 20138 Milan, Italy.

Riassunto

Background

The aim of this study was to assess the accuracy of a comprehensive multidetector computed tomography (MDCT) evaluation of the aortic annulus (AoA), coronary artery disease (CAD), and peripheral vessels in patients referred for transcatheter aortic valve implantation (TAVI).

Methods

In 60 patients referred for TAVI, the following parameters were assessed with 64-slices MDCT and compared with transesophageal echocardiography (TEE), invasive coronary angiography (ICA), and peripheral angiography: AoA maximum diameter (Max-AoA-DMDCT), minimum diameter (Min-AoA-DMDCT), and area; lumen morphology index ([Max-AoA-DMDCT/Min-AoA-DMDCT]); length of the left, right, and non-coronary aortic leaflets; degree (grades 1-4) of aortic leaflet calcifications; distance between AoA and left main coronary ostium and between AoA and right coronary ostium CAD and peripheral vessel disease.

Results

The Max-AoA-DMDCT and Min-AoA-DMDCT were 25.1 ± 2.8 and 21.2 ± 2.2 mm, respectively, with high correlation versus AoA diameter measured with TEE (r = 0.82 and 0.86, respectively). The area of AoA, systolic and diastolic lumen morphology index were 410 ± 81.5 mm2, 1.19 ± 0.1 and 1.22 ± 0.11, respectively. Aortic leaflet calcification score was 3.3 ± 0.5. The lengths of left, right, and non-coronary aortic leaflets were 14.2 ± 2.4, 13.7.1 ± 2.1, and 14.5 ± 2.6 mm, whereas distances between AoA and the left main coronary ostium and between AoA, and the right coronary ostium were 13.7 ± 2.9 and 15.8 ± 3.5 mm, respectively. Feasibility, negative predictive value, and accuracy for CAD detection versus ICA were 87%, 100% (CI 100-100), and 96% (95% CI 94-100), respectively. All patients (N = 17) who were ineligible for TAVI were correctly detected by MDCT.

Conclusions

A comprehensive MDCT evaluation of patients referred for TAVI is feasible, provides more accurate assessment than TEE of AoA morphology, and may replace peripheral angiography in all patients and ICA in patients without significant CAD.

Il testo completo di questo articolo è disponibile in PDF.

Abbreviations : AoA, AoA-A, AoA-AMDCT, AoA-D, AoA-DTEE, AoAMDCT, AoAMDCT-LM, AoAMDCT-RCA, ALC, AS, AVR, CABG, CAD, ECG, Fe, HR, ICA, LCL, LMI, Max-AoA-DMDCT, Min-AoA-DMDCT, MDCT, NPV, PPV, PVD, RCL, Se, Sp, TAVI, TEE, TTE


Mappa


© 2011  Mosby, Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 161 - N° 6

P. 1106-1113 - Giugno 2011 Ritorno al numero
Articolo precedente Articolo precedente
  • Safety, feasibility, and diagnostic value of cardiac magnetic resonance imaging in patients with cardiac pacemakers and implantable cardioverters/defibrillators at 1.5 T
  • Claas P. Naehle, Jens Kreuz, Katharina Strach, Jörg Otto Schwab, Simon Pingel, Roger Luechinger, Rolf Fimmers, Hans Schild, Daniel Thomas
| Articolo seguente Articolo seguente
  • Clinical outcome and predictors for adverse events after transcatheter aortic valve implantation with the use of different devices and access routes
  • Peter Wenaweser, Thomas Pilgrim, Nadja Roth, Alexander Kadner, Stefan Stortecky, Bindu Kalesan, Fabienne Meuli, Lutz Büllesfeld, Ahmed A. Khattab, Christoph Huber, Balthasar Eberle, Gabor Erdös, Bernhard Meier, Peter Jüni, Thierry Carrel, Stephan Windecker

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.