Safety, feasibility, and diagnostic value of cardiac magnetic resonance imaging in patients with cardiac pacemakers and implantable cardioverters/defibrillators at 1.5 T - 06/08/11

Doi : 10.1016/j.ahj.2011.03.007 
Claas P. Naehle, MD a, , Jens Kreuz b, Katharina Strach a, Jörg Otto Schwab b, Simon Pingel b, Roger Luechinger, PhD c, Rolf Fimmers, PhD d, Hans Schild, MD a, Daniel Thomas, MD a
a Department of Radiology, University of Bonn, Bonn, Germany 
b Department of Medicine - Cardiology, University of Bonn, Bonn, Germany 
c Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland 
d Department of Medical Biometry, Informatics, and Epidemiology, University of Bonn, Bonn, Germany 

Reprint requests: Claas P. Naehle, MD, Department of Radiology, University of Bonn, Sigmund-Freud-Str. 25, 53105 Bonn, Germany.

Riassunto

Background

Recent studies suggest that magnetic resonance (MR) imaging of the brain and spine may safely be performed in patients with pacemakers (PMs) and implantable cardioverter/defibrillators (ICDs), when taking adequate precautions. The aim of this study was to investigate safety, feasibility, and diagnostic value (DV) of MR imaging in cardiac applications (cardiac MR [CMR]) in patients with PMs and ICDs for the first time.

Methods

Thirty-two PM/ICD patients with a clinical need for CMR were examined. The specific absorption rate was limited to 1.5 W/kg. Devices were reprogrammed pre-CMR to minimize interference with the electromagnetic fields. Devices were interrogated pre-CMR and post-CMR and after 3 months. Troponin I levels were measured pre-CMR and post-CMR; image quality (IQ) and DV of CMR were assessed.

Results

All devices could be reprogrammed normally post-CMR. No significant changes of pacing capture threshold, lead impedance, and troponin I were observed. Image quality in patients with right-sided devices (RSD) was better compared with that in patients with left-sided devices (LSD) (P < .05), and less myocardial segments were affected by device-related artefacts (P < .05). Diagnostic value was rated as sufficiently high, allowing for diagnosis, or better in 12 (100%) of 12 patients with RSD, and only in 7 (35%) of 20 patients with LSD.

Conclusions

Cardiac MR may be performed safely when limiting specific absorption rate, appropriately monitoring patients, and following device reprogramming. Cardiac MR delivers good IQ and DV in patients with RSD. Cardiac MR in patients with RSD may therefore be performed with an acceptable risk/benefit ratio, whereas the risk/benefit ratio is rather unfavorable in patients with LSD.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2011  Mosby, Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 161 - N° 6

P. 1096-1105 - Giugno 2011 Ritorno al numero
Articolo precedente Articolo precedente
  • Patterns of left ventricular remodeling in chronic heart failure: Prevalence and prognostic implications
  • Frank Lloyd Dini, Paola Capozza, Francesca Donati, Anca Simioniuc, Anca Irina Corciu, Paolo Fontanive, Andrea Pieroni, Vitantonio Di Bello, Mario Marzilli
| Articolo seguente Articolo seguente
  • Feasibility and accuracy of a comprehensive multidetector computed tomography acquisition for patients referred for balloon-expandable transcatheter aortic valve implantation
  • Gianluca Pontone, Daniele Andreini, Antonio L. Bartorelli, Andrea Annoni, Saima Mushtaq, Erika Bertella, Alberto Formenti, Sarah Cortinovis, Francesco Alamanni, Melissa Fusari, Veronica Bona, Gloria Tamborini, Manuela Muratori, Giovanni Ballerini, Cesare Fiorentini, Paolo Biglioli, Mauro Pepi

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2024 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.