Comparaison de la mesure de la PETCO2 obtenue avec le Smart Capnoline™ avec la PaCO2 en postopératoire de chirurgie cardiaque chez des patients intubés puis extubés - 09/02/11

Doi : 10.1016/j.annfar.2010.11.008 
S. Rousselon a, M. Coat a, B.V. Nguyen a, P. Gouny b, E. Nowak c, J.-P. Wargnier a, C.-C. Arvieux a, G. Gueret a,
a Département d’anesthésie réanimation, CHU La Cavale-Blanche, boulevard Tanguy-Prigent, 29609 Brest, France 
b Service de chirurgie cardiaque, thoracique et vasculaire, CHU La Cavale-Blanche, boulevard Tanguy-Prigent, 29609 Brest, France 
c Inserm CIC 0502, CHU La Cavale-Blanche, boulevard Tanguy-Prigent, 29609 Brest, France 

Auteur correspondant.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

pagine 4
Iconografia 4
Video 0
Altro 0

Résumé

Objectifs

Comparer la PETCO2 obtenue avec le Smart Capnoline™ et la PaCO2 en postopératoire de chirurgie cardiaque.

Type d’étude

Prospective observationnelle.

Patients

Vingt patients après chirurgie cardiaque.

Méthodes

Aux soins intensifs, lorsqu’une gazométrie artérielle était prélevée, la valeur de la PETCO2 était relevée. Trois capteurs ont été utilisés : Filterline H set pour les patients intubés, Smart Capnoline HO2 (version nasale) et Smart Capnoline O2 (version bucconasale) après extubation. Les données ont été comparées par un test de Wilcoxon et le calcul du coefficient de corrélation intra-classe.

Résultats

La différence PaCO2 – PETCO2 augmente après extubation, que ce soit avec le capteur bucconasal ou avec le capteur nasal (patients intubés : 6,6±4,3mmHg, capteur nasal : 9,3±3,5mmHg, capteur bucconasal : 15,4±12,9mmHg). La différence plus importante observée avec le capteur bucconasal disparaît si on exclut un patient ayant nécessité une ré-intubation pour détresse respiratoire.

Conclusion

En postopératoire de chirurgie cardiaque, si la mesure de la PETCO2 permet une estimation satisfaisante de la PaCO2 chez les patients intubés, cette estimation est moins fiable après extubation. Le capteur bucconasal ne semble pas apporter d’amélioration par rapport au capteur nasal après chirurgie cardiaque.

Il testo completo di questo articolo è disponibile in PDF.

Abstract

Objective

To compare the PaCO2 with the ETCO2 obtained with the Smart Capnoline™ in the postoperative setting of cardiac surgery during ventilation and after extubation

Type of study

Prospective, observational.

Patients

Twenty patients after cardiac surgery.

Methods

In the intensive care unit, arterial blood gases were measured concomitantly with ETCO2, and difference between PaCO2 and ETCO2 were calculated. Three CO2 sensors were utilized: Filterline H set for intubated patients, Smart Capnoline HO2 (nasal version) and Smart Capnoline O2 (bucconasal version) after extubation. Data were compared with Wilconson test and the intraclass correlation coefficient was calculated.

Results

The difference PaCO2 – ETCO2 was significantly larger in extubated patients compared to intubated patients, which is also confirmed for the bucconasal sensor (intubated patients: 6.6±4.3mmHg, nasal sensor: 9.3±3.5mmHg, bucconasal sensor: 15,4±12.9mmHg).

Conclusion

In the postoperative setting of cardiac surgery, ETCO2 measurements allow a reliable estimation of PaCO2 in intubated patients in contrast to measurements in extubated patients. The bucconasal CO2 sensor does not show more reliable measurements compared to nasal sensors in the postoperative setting of cardiac surgery.

Il testo completo di questo articolo è disponibile in PDF.

Mots clés : PETCO2, Chirurgie cardiaque

Keywords : End-tidal CO2, Cardiac surgery


Mappa


© 2010  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 30 - N° 1

P. 13-16 - Gennaio 2011 Ritorno al numero
Articolo precedente Articolo precedente
  • Analyse de concordance entre l’échographie pulmonaire et la radiographie thoracique en réanimation
  • M. Gazon, N. Eboumbou, M.-O. Robert, P. Branche, S. Duperret, J.-P. Viale
| Articolo seguente Articolo seguente
  • L’acide tranexamique diminue les hématomes mais pas les douleurs après prothèse totale de genou
  • I. Chevet, F. Remérand, C. Couvret, A. Baud, C. Pouplard, P. Rosset, M. Laffon, J. Fusciardi

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2024 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.