Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw - 16/02/10
, U. Yavuz b, B. Kadıoğlu c, B. Akman d, V. Kılınçoğlu e, K. Ünay c, F. Altıntaş f| pagine | 2 |
| Iconografia | 0 |
| Video | 0 |
| Altro | 0 |
Summary |
Background |
Tip-apex distance greater than 25mm is accepted as a strong predictor of screw cut-out in patients with intertrochanteric femoral fracture treated by dynamic hip screw. The aim of this retrospective study was to evaluate the position of the screw in the femoral head and its effect on cut-out failure especially in patients with inconvenient tip-apex distance.
Materials and methods |
Sixty-five patients (42 males, 23 females; mean age of 57.6 years) operated by dynamic hip screw for intertrochanteric femoral fractures were divided in two groups taking into consideration the tip-apex distance less (group A; 14 patients) or more (group B; 51 patients) than 25mm. Patient’s age and gender, follow-up period, fracture type, degree of osteoporosis, reduction quality of the fracture, position of the screw in the femoral head, number of patients with cut-out failure and Harris hip score were compared.
Results |
The average follow-up time was 41.7 months. The mean tip-apex distance was 17.14mm in group A and 36.67mm in group B. One (7.1%) patient in group A and three (5.8%) patients in group B had screw cut-out. Except the screw position, no statistical differences were observed between the two groups with regards to study data. The screw was placed in femoral head more inferiorly (p=0.045) on frontal and more posteriorly (p=0.013) on sagital planes in group B, while central placement of the screw was present in group A. The common characteristics of three patients with screw cut-out in group B was the position of the screw which was located in femoral head more superiorly and anteriorly after an acceptable fracture reduction.
Conclusions |
Peripheral placement of the screw in femoral head increases tip-apex distance. However posterior and inferior locations may help to support posteromedial cortex and calcar femoral in unstable intertrochanteric fractures and reduce the risk of cut-out failure.
Level of evidence |
IV, retrospective series.
Il testo completo di questo articolo è disponibile in PDF.Keywords : Intertrochanteric fracture, Dynamic hip screw, Screw position, Tip-apex distance, Cut-out failure
Mappa
| Cet article peut être consulté in extenso dans la version anglaise de la revue Orthopaedics & Traumatology: Surgery & Research sur Science Direct (www.sciencedirect.com/) en utilisant le DOI ci-dessus. |
Vol 96 - N° 1
P. 33-34 - febbraio 2010 Ritorno al numeroBenvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.
Già abbonato a @@106933@@ rivista ?