Validation du logiciel MediCAD® pour l’aide à la planification préopératoire entièrement digitale des prothèses totales de hanche : une étude rétrospective - 21/11/24

Validation of MediCAD® software for fully digital preoperative planning of total hip arthroplasty: A retrospective study

Doi : 10.1016/j.rcot.2024.07.010 
Pierre-Alban Bouché 1, , Simon Corsia 2, Guillaume Auberger 3, Jules Descamps 1, Philippe Anract 2, Moussa Hamadouche 2
1 Hôpital Lariboisière, service de chirurgie orthopédique et traumatologique, 2, rue Ambroise-Paré, 75010 Paris, France 
2 Hôpital Cochin, service de chirurgie orthopédique et traumatologique, 27, rue du Faubourg-Saint-Jacques, 75014 Paris, France 
3 Groupe hospitalier Diaconesses Croix St-Simon, service de chirurgie orthopédique et traumatologique, 125, rue d’Avron, 75020 Paris, France 

Pierre-Alban Bouché, Hôpital Lariboisière, service de chirurgie orthopédique et traumatologique, 2, rue Ambroise-Paré, 75010 Paris, France.Hôpital Lariboisière, service de chirurgie orthopédique et traumatologique2, rue Ambroise-ParéParis75010France

Résumé

Introduction

L’étape de planification qui précède l’implantation d’une prothèse totale de hanche (PTH) reste indispensable. Les logiciels de planification numérique sont de plus en plus répandus et peu d’études rapportent des résultats sur la reproductibilité de ces outils, et aucun en France sur la suite logicielle MediCAD® pourtant largement diffusée. Aussi nous avons mené une étude rétrospective afin : 1) d’évaluer la précision et la fiabilité de ce logiciel de planification, 2) évaluer la reproduction intra- et interobservateur, 3) mesurer l’influence de l’obésité sur précision de la planification.

Hypothèse

La planification d’une PTH à l’aide de la suite MediCAD® était fiable et reproductible.

Patients et méthodes

Il s’agit d’une étude de cohorte rétrospective monocentrique. Cent-un cas consécutifs de PTH posées par un chirurgien senior expérimenté ont été planifiés à deux reprises par 2 chirurgiens maintenus en aveugle. L’implant acétabulaire était cimenté dans 90 cas (79,2 %) et non-cimenté dans 11 cas (10,9 %). L’implant acétabulaire était à double mobilité dans 21 cas (20,8 %). La tige fémorale était cimentée dans 60 cas (59,4 %). Le critère de jugement était le nombre de planifications jugées exactes (même taille que l’implant posé). La reproductibilité intra-observateur et interobservateur ont été calculées à l’aide du coefficient de corrélation intraclasse et interclasse (CIC) avec l’intervalle de confiance à 95 % (IC95 %).

Résultats

Une concordance exacte (aucune différence) a été obtenue pour 15 planifications (14,9 %) et pour 45 planifications (44,6 %) respectivement pour l’implant acétabulaire et fémoral par rapport à l’implant posé pour le premier investigateur. Le second investigateur a obtenu une concordance exacte (aucune différence) pour 20 planifications (19,8 %) et pour 50 planifications (49,5 %) respectivement pour l’implant acétabulaire et fémoral. La reproductibilité intra-observateur pour l’implant acétabulaire était jugée moyenne (CIC=0,57 ; IC95 % [0,43–0,69]) et faible (CIC=0,38 ; IC95 % [0,20–0,54]) respectivement pour le 1er et le 2nd investigateur. La reproductibilité intra-observateur pour la tige fémorale était jugée faible pour le 1er (CIC=0,47 ; IC95 % [0,30–0,61]) et le 2nd investigateur (CIC=0,45 ; IC95 % [0,29–0,60]). La reproductibilité interobservateur était jugée faible entre les deux investigateurs concernant la planification des implants acétabulaire (CIC=0,39 ; IC95 % [0,21–0,54]) et fémoral (CIC=0,42 ; IC95 % [0,24–0,57]). Au total, en regroupant les deux investigateurs, une prévision exacte (aucune différence avec l’implant définitif) de la taille de l’implant acétabulaire était obtenue dans 31 cas (18,9 %) chez les patients non obèses, et dans 7 cas (20,6 %) chez les patients obèses (p=0,62).

Discussion

Les résultats affichent une fiabilité acceptable du logiciel MediCAD®. L’apport de l’expérience et de l’agrandissement radiographique et l’hétérogénéité dans les implants (cimentés vs non cimentés) influencent le résultat de la planification mais pas l’obésité dans cette étude. Les moyens actuels ne permettent pas encore en planification bidimensionnelle, d’obtenir une échelle radiologique systématiquement fiable. Le scanner est plébiscité par certains, mais son coût économique est supérieur et il présente toujours l’inconvénient d’une irradiation majorée. Cette étude montre que le logiciel MediCAD® peut être utilisé avec des résultats satisfaisants pour la planification préopératoire d’une prothèse totale de hanche.

Niveau de preuve

III ; série rétrospective comparative diagnostique.

Il testo completo di questo articolo è disponibile in PDF.

Summary

Introduction

The planning step that precedes a total hip arthroplasty (THA) procedure is crucial. Digital planning software programs are being increasingly used, although few studies have reported on the reliability of such tools. Furthermore, no studies have been conducted on the MediCAD® software, despite it being widely used in France. This led us to conduct a retrospective study to: 1) assess the accuracy of this planning software, 2) determine the intra- and inter-rater reliability, 3) determine how obesity affects the accuracy of planning.

Hypothesis

THA planning is accurate and reliable when using the mediCAD® software.

Patients and methods

This was a single center, retrospective study. One hundred one consecutive cases performed by a single experienced surgeon were planned retrospectively by two blinded surgeons on two separate occasions. The acetabular cup was cemented in 90 hips (89%), cementless in 11 hips (11%). A dual mobility cup was used in 21 hips (21%). The femoral stem was cemented in 60 hips (59%). The endpoint was the number of exact plans, defined as the same size as the actual implants. An acceptable match was defined as a difference of one size. The match was unacceptable if the planned and implanted size differed by more than 2 for the acetabular cup or by more than 1 size for the femoral stem. The intra-rater and inter-rater reliability were calculated using the intraclass correlation coefficient (ICC) with 95% confidence intervals (CI).

Results

Exact agreement was found by the first rater for 15 planned acetabular cups (15%) and for 45 planned femoral stems (45%) relative to the implants used. The second rater reached exact agreement for 20 planned acetabular cups (20%) and 50 planned femoral stems (50%). The intra-rater reliability for the acetabular cup was average (ICC=0.57; 95%CI [0.43–0.69]) and poor (ICC=0.38; 95%CI [0.20–054]) for the 1st and 2nd rater, respectively. The intra-rater reliability for the femoral stem was poor for the 1st rater (ICC=0.47 95%CI [0.30–0.61]) and the 2nd rater (ICC=0.45; 95%CI [0.29–0.60]). The interobserver reliability was low for the planned acetabular cup (ICC=0.39; 95%CI [0.21–0.54]) and the planned femoral stem (ICC=0.42; 95%CI [0.24–0.57]). Overall, when combining the two raters, exact prediction of the acetabular cup was achieved in 31 hips (19%) in non-obese patients and in 7 hips (21%) in obese patients (P=0.62).

Discussion

This study found acceptable reliability of the mediCAD® software. Experience level, radiograph magnification affected the planning outcome in this study, but obesity did not. We currently do not have the ability to incorporate a reliable radiological scale for two-dimensional templating. Some surgeons prefer using a CT scan, but this costs more than conventional radiographs and exposes the patient to more radiation. This study shows that the MediCAD® software can provide satisfactory output for the preoperative planning of THA.

Level of evidence

III; retrospective, diagnostic, comparative study.

Il testo completo di questo articolo è disponibile in PDF.

Mots clés : Arthroplastie totale de hanche, Modèle préopératoire, Modélisation numérique

Keywords : Total hip arthroplasty, Preoperative templating, Digital templating


Mappa


 Ne pas utiliser, pour citation, la référence française de cet article, mais celle de l’article original paru dans Orthopaedics & Traumatology: Surgery & Research, en utilisant le DOI ci-dessus.


© 2024  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 110 - N° 8

P. 1142-1149 - Dicembre 2024 Ritorno al numero
Articolo precedente Articolo precedente
  • Un protocole scanner simple dédié à la planification de prothèse totale de hanche pour fracture du col du fémur
  • Jean Langlois, Thais Dutra Vieira, Tarik Ait Si Selmi, Michel P. Bonnin
| Articolo seguente Articolo seguente
  • Simple Hip Value : un score simple pour l’évaluation de la hanche
  • Alix Addi, Tristan Duguay, Eugenie Valentin, Philippe Anract, Alexandre Hardy

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2024 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.