Révision isolée de la cupule acétabulaire par chirurgie assistée par ordinateur - 07/11/19

Isolated acetabular revision with femoral stem retention using computed tomography-based navigation

Doi : 10.1016/j.rcot.2019.09.002 
Yuta Kubota a, Nobuhiro Kaku a, , Hiroaki Tagomori a, Masashi Kataoka b, Hiroshi Tsumura a
a Department of Orthopaedic Surgery, Faculty of Medicine, Oita University 1-1 Idaigaoka, Hasama-machi Yufu City, 879-5593 Oita, Japon 
b Physical Therapy Course of Study, Faculty of Welfare and Health Sciences, Oita University 700 Dannoharu, Oita City, 870-1192 Oita, Japon 

Auteur correspondant.

Abstract

Background

In isolated acetabular revision surgery, surgeons must place the cup at an appropriate angle with various retained stem anteversion angles to prevent postoperative dislocation. For accurate acetabular cup position, various navigation systems have been used. Nevertheless, no publications have reported combined cup and stem anteversion and dislocation rates after isolated acetabular revision, especially comparing the use of navigation with manual implantation. Therefore we performed a retrospective comparative study to answer the following questions: 1) What is the combined anteversion after isolated acetabular revision with computed tomography-based navigation? 2) Does navigation improve the accuracy of cup angle and combined anteversion, 3) reduce dislocation rate, and 4) reduce operative time?

Hypothesis

A navigation system makes combined anteversion near the target angle in isolated acetabular revision.

Patients and methods

We conducted a retrospective study of 32 hips in 24 patients who underwent isolated acetabular revision total hip arthroplasty using computed tomography-based navigation system. The control group comprised 8 hips in 8 patients who underwent the same procedure without navigation.

Results

In the navigation group, average Widmer's combined anteversion was 39.0±8.7 (range, 25.3–56.6). Cup positions were 40.3±2.9 (range, 29.0–49.0) for radiographic abduction angle, 24.3±8.0 (range, 4.6–42.6) for radiographic anteversion. In the control group, Widmer's combined anteversion was 47.2±15.1 (range, 27.0–74.3, p=0.048). The average cup positions for radiographic abduction angle and anteversion were 36.7±8.9 (range, 24.5–54.9) and 29.1±7.3 (range, 17.2–38.8), respectively (p>0.05). Widmer's combined anteversion deviated from the target angle by a smaller amount in the navigation group than in the control group: errors in measurement of Widmer's combined anteversion were 7.2±5.1 (range, 0.74–19.6) in the navigation group and 13.9±11.1 (range, 3.6–37.3) in the control group (p=0.135). Postoperative dislocation occurred in none of the 32 hips (0%) in the navigation group and in one of the 8 hips (12.5%) in the control group (p=0.2).

Discussion

Using the navigation system, combined anteversion is made near the target angle in isolated acetabular revision surgery and more accurately than manual implantation. These results should be considered as preliminary since this is a limited cohort, but it brings new knowledge in navigation considering the very limited number of series using of navigation in isolated cup revision of total hip arthroplasty. In addition, this is the first study to investigate combined cup and stem anteversion in isolated acetabular revision. The use of CT-scan is helpful to diagnose error in stem anteversion and to adapt the orientation of the new cup.

Level of evidence

III, retrospective case control study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Total hip arthroplasty, X-ray computed tomography, Bone anteversion, Acetabulum, Retrospective study



 Cet article peut être consulté in extenso dans la version anglaise de la revue Orthopaedics & Traumatology: Surgery & Research sur Science Direct (sciencedirect.com) en utilisant le DOI ci-dessus.


© 2019  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 105 - N° 7

P. 860 - Novembre 2019 Ritorno al numero
Articolo precedente Articolo precedente
  • Quels sont les facteurs de risque de luxation après révision de prothèse totale de hanche à l’aide d’une cupule à double mobilité ? Une étude cas-témoins à propos de 16 cas
  • Denis Huten, Yohann Fournier, Thomas Gicquel, Pierre Bertho, Frédéric Basselot, Moussa Hamadouche
| Articolo seguente Articolo seguente
  • Résultats cliniques et dynamométriques de la réparation du système abducteur de hanche par plaque d’hydroxyapatite trochantérienne sur une prothèse modulaire après résection tumorale du fémur proximal
  • Vincent Crenn, Sylvain Briand, Philippe Rosset, Jean-Camille Mattei, Alban Fouasson-Chailloux, Louis-Romée Le Nail, Denis Waast, Mickael Ropars, François Gouin

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2024 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.