Accuracy of joint-preserving tumor resections performed by navigation or patient-specific instrument. A cadaveric comparative study - 23/11/17

Précision des résections tumorales conservatrices réalisées par navigation ou instrument sur mesure. Étude comparative sur cadavres

Doi : 10.1016/j.rcot.2017.09.266 
Laurent Paul ⁎ , Sarah Bosma, Kwok Chuen Wong, Jasper Gerbers, Paul Jutte
 Woluwé-Saint-Lambert, Belgium 

⁎Corresponding author.

Riassunto

Introduction

Computer assisted surgery (CAS) and patient specific instrument (PSI) have been reported to increase accuracy and predictability of tumor resections. The challenging joint-preserving surgery that retains the native joint with an optimal function may benefit from these assistances. Literature has shown that the pelvic surgery can be improved by these assistances, but to date, there is no clear statement around the knee. This cadaver study aims at investigating the surgical accuracy of CAS and PSI in joint-preserving surgery of knee joint.

Matériel et méthodes

CT-scans of four cadavers (8 legs) were acquired. Eight metaphyseal bone sarcomas were simulated on distal femur and proximal tibia. Joint-preserving multiplanar resections were planned. Four techniques (freehand, CAS, PSI and CAS+PSI) were used to perform the resections. Total resection time was measured for each. Postoperative CT-scans were acquired and analysed to measure the location accuracy (maximum deviation of distance between the planned and achieved resections).

Résultats

Both CAS+PSI and PSI techniques could reproduce the planned resections with mean location accuracy below 2mm, compared to 3.6mm for CAS and 9.2mm for the freehand technique. No statistical difference has been found between the CAS+PSI and PSI techniques (P=0.92). CAS and CAS+PSI were found significantly more accurate than CAS (respectively P=0.042 and P=0.034). Finally, freehand technique was found significantly less accurate than any other technique (P<0.05). PSI was significantly fastest (mean 4.8min) when compared to CAS+PSI technique (mean 12.8min, P<0.001) and CAS technique (mean 17.0min, P=<0.001). No difference was found between PSI and freehand technique (mean 6.8min).

Discussion

This cadaveric study of simulated joint-preserving tumor surgery around the knee joint suggests that CAS, PSI or CAS+PSI techniques permitted to reproduce the multiplanar resections as planned. PSI technique could achieve the most accurate bone resections.

Conclusion

The accuracy obtained by using PSI, CAS or PSI+CAS is clinically sufficient to ensure a safe margin. However, the PSI accuracy depends on its stability on the bone surface. This is directly relative to the PSI design and the communication between surgeon and engineering team to define surgical approach and available bone surface.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2017  Pubblicato da Elsevier Masson SAS.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 103 - N° 7S

P. S137 - novembre 2017 Ritorno al numero
Articolo precedente Articolo precedente
  • Modélisation biomécanique 3D de la cinématique du pied sous activation musculaire avec et sans arthrodèse de cheville
  • Antoine Perrier, Wilfrid Graff, Marek Bucki, Yohan Payan
| Articolo seguente Articolo seguente
  • How significant is statistically significant? A literature review
  • Ismat Ghanem, Nour Khalil, Aren Joe Bizdikian, Fares Yared, Ayman Assi

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.