Analyse de coût comparant les fibroscopes à usage unique (Ambu® aScope™) et les fibroscopes réutilisables pour l’intubation difficile - 06/06/13

Doi : 10.1016/j.annfar.2013.01.014 
M. Aïssou a, M. Coroir a, C. Debes a, T. Camus a, N. Hadri b, C. Gutton a, M. Beaussier a,
a Département d’anesthésie-réanimation chirurgicale, université Pierre et Marie-Curie Paris 6, hôpital Saint-Antoine, Assistance publique-Hôpitaux de Paris (AP–HP), 184, rue du Faubourg-Saint-Antoine, 75571 Paris cedex 12, France 
b Service de pharmacie, université Pierre et Marie-Curie Paris 6, hôpital Saint-Antoine, Assistance publique-Hôpitaux de Paris (AP–HP), 184, rue du Faubourg-Saint-Antoine, 75571 Paris cedex 12, France 

Auteur correspondant.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

páginas 5
Iconografías 2
Vídeos 0
Otros 0

Résumé

Objectifs

Les recommandations incitent chaque bloc opératoire à disposer d’un fibroscope pour l’intubation trachéale difficile. Le matériel réutilisable pose le problème de sa disponibilité, de sa maintenance, du risque infectieux et de son coût d’acquisition. Des fibroscopes à usage unique sont désormais disponibles (Ambu® aScope™). Cette étude a pour objectif d’évaluer le coût de la fibroscopie utilisant un matériel réutilisable et de le comparer à celui de l’usage unique.

Type d’étude

Étude de type minimisation de coût.

Matériel étudié

Fibroscope réutilisable Pentax® FB 15P (Pentax France, Argenteuil) et fibroscope à usage unique Ambu® aScope™ (Ambu A/S, Ballerup, Danemark).

Méthode

Évaluation du coût amorti par utilisation de deux fibroscopes réutilisables incluant le prix d’achat, les dépenses de maintenance et de désinfection sur une période de six années. Le coût d’utilisation du fibroscope Ambu® aScope™ correspond à son prix d’achat.

Résultats

Le coût des fibroscopes réutilisables a été de 55 874€ pour 270 intubations, soit un coût unitaire de 206€ par fibroscopie. Durant cette période, 780 procédures de nettoyage-désinfection ont été réalisées pour un coût de 32 611€. Les dépenses d’investissement et de maintenance ont été respectivement de 18 382€ et de 4880€. Ce coût est à rapporter à celui du fibroscope à usage unique, de 200€ par fibroscopie.

Conclusion

Sur une période de six ans, les coûts liés à l’utilisation des fibroscopes réutilisables et à usages uniques sont proches. Les avantages de l’usage unique conduisent à discuter l’intérêt de ce matériel pour l’intubation difficile prévue dans les blocs opératoires.

El texto completo de este artículo está disponible en PDF.

Abstract

Objective

Recommendations on difficult tracheal intubation constrain each surgical block to have a fiberscope available. Reusable fiberscope presents the problem of availability, the risk of non-conventional infections transmission and the cost. Single-use fiberscopes are presently available (Ambu® aScope™). This medico-economic study aims to assess the cost of using a reusable fiberscope as compared to the single use fiberscope.

Study design

Cost minimization analysis.

Material

Reusable fiberscope Pentax® FB 15P (Pentax France, Argenteuil) and single-use fiberscope Ambu® aScope™ (Ambu A/S, Ballerup, Denmark).

Method

Minimization-cost analysis conducted between 2006 and 2012. The amortization cost per utilization for two reusable fiberscopes took into account the acquisition and maintenance costs, as well as the costs related to disinfection. The cost of the single use fiberscope was calculated according to its acquisition cost.

Results

The total cost of the reusable material was 55,874€ over 6years, corresponding to a unitary cost of 206€ per fiberscopy. During this period, 780 sterilizations were carried out for a total cost of 32,611€. Acquisition and maintenance costs were respectively 18,382€ and 4880€. The cost of the single use fiberscope is of 200€ per fiberscopy.

Conclusion

This medico-economic evaluation shows that the utilization cost of single use and reusable fiberscopes are very close. This should be analyzed at the light of some benefits of using single use devices for the difficult tracheal intubation.

El texto completo de este artículo está disponible en PDF.

Mots clés : Anesthésie, Fibroscope, Évaluation médico-économique, Intubation difficile

Keywords : Anesthesia, Fiberscope, Medico-economic analysis, Difficult tracheal intubation


Esquema


© 2013  Société française d’anesthésie et de réanimation (Sfar). Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 32 - N° 5

P. 291-295 - mai 2013 Regresar al número
Artículo precedente Artículo precedente
  • Revue des erreurs médicamenteuses (REMED) : une première en service de réanimation
  • M. Guillaudin, B. Debien, F. Aouadene, G. Camus, O. Galvez, M. Paillet, X. Bohand
| Artículo siguiente Artículo siguiente
  • Comparaison de l’Airtraq™ et du laryngoscope Macintosh pour l’intubation trachéale chez l’adulte avec stabilisation manuelle en ligne du rachis cervical
  • M. Amor, S. Nabil, M. Bensghir, A. Moussaoui, S. Kabbaj, N. Drissi Kamili, W. Maazouzi

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
La compra de artículos no está disponible en este momento.

¿Ya suscrito a @@106933@@ revista ?

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.