Poor short-term outcomes after computer-assisted rotating-platform total knee arthroplasty with a deep-trochlear-groove femoral component: Analysis of 19 patients - 30/03/13

Doi : 10.1016/j.otsr.2012.07.010 
A. Akakpo a, , b , G. Dereudre a, b, N. Fouilleron a, b, G. Pasquier a, b, H. Migaud a, b
a Nothern France Lille University, 59000 Lille, France 
b Department of Orthopaedics, Salengro Hospital, Lille Regional Teaching Hospital Center, place de Verdun, 59037 Lille cedex, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Summary

Background

Congruent rotating tibial plateaus are designed to minimise wear after total knee arthroplasty (TKA). The Score™ prosthesis has a congruent rotating tibial plateau, a deep trochlear groove, and uses a computer-assisted navigation system for ligament balancing. Although this prosthesis is widely utilized, no accurate data on outcomes are available.

Hypothesis

The innovative features of the Score™ prosthesis, most notably patellar replacement with a highly constrained femoral component, do not jeopardize implant survival.

Patients and methods

In a pilot study, we retrospectively evaluated outcomes of 19 patients treated with Score™ knee replacement between February and October 2006 (mean age, 66.8years; range, 58–82years). The evaluation criteria were the International Knee Society (IKS) scores and prosthesis survival rate estimated using Kaplan-Meier plots with failure defined as revision need to change the prosthesis.

Results

Mean follow-up was 35.3months. The IKS knee score increased from 27.4 (5–60) preoperatively to 81.4 (45–99) at last follow-up (P<0.0001). Mean mechanical axis was 181.2° (180–186°), with 16 between 180° and 183°. Revision surgery was required in five cases (for patellar complications with combined motion-range limitation in flexion (<90°) and extension (5–20°) in three cases, isolated motion-range limitation in one case, and recurvatum deformity with instability in one case). Prosthesis survival was 82% (73–91%) after 24months and 65% (51–78%) after 44months.

Discussion

The deep trochlear groove femoral component resulted in patellar complications, which were the most common reasons for revision surgery, together with motion-range limitation and instability possibly related to improper use of the navigation system. This small retrospective case-series study showed an unusually low prosthesis survival rate probably related to the implant design. We no longer use the Score™ prosthesis, despite the availability of a dedicated navigation system, and we recommend careful monitoring of patients who have this prosthesis.

Level of evidence

Level IV, retrospective study.

El texto completo de este artículo está disponible en PDF.

Keywords : Knee prosthesis, Rotating platform, Computer-assisted implant surgery, Patello-femoral, Instability


Esquema


© 2012  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 99 - N° 2

P. 183-190 - avril 2013 Regresar al número
Artículo precedente Artículo precedente
  • Unstable pelvic ring injury with hemodynamic instability: What seems the best procedure choice and sequence in the initial management?
  • S. Abrassart, R. Stern, R. Peter
| Artículo siguiente Artículo siguiente
  • Simultaneous bilateral total knee arthroplasty. A multicenter feasibility study
  • J.-Y. Jenny, C. Trojani, J.-L. Prudhon, C. Vielpeau, D. Saragaglia, C. Houillon, T. Ameline, F. Steffan, B. Bugnas, J. Arndt, Hip and Knee Surgery French Society (SFHG)

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2025 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.