Suscribirse

Endoscopic management of GI perforations with a new over-the-scope clip device (with videos) - 24/08/11

Doi : 10.1016/j.gie.2010.04.006 
Andrea Parodi, MD, Alessandro Repici, MD, Antonietta Pedroni, MD, Sabrina Blanchi, MS, Massimo Conio, MD
Current affiliations: Department of Gastroenterology (A. Parodi, A. Pedroni, S.B., M.C.), General Hospital, San Remo; Digestive Endoscopy Unit (A.R.), IRCCS, Istituto Clinico Humanitas, Rozzano, Milan, Italy 

Reprint requests: Massimo Conio, MD, Director, Department of Gastroenterology, General Hospital, 18038, San Remo, Italy

Resumen

Background

Through-the-scope endoclips have been used to manage small perforations in the GI tract, but they have limitations. A new over-the-scope clip system, OTSC (Ovesco Endoscopy, Tuebingen, Germany), may be suitable for the closure of larger GI leaks.

Objective

To evaluate the clinical outcomes of patients with GI perforations of up to 20 mm, treated with OTSC.

Design

Prospective, single-arm, pilot study.

Setting

General hospitals referral centers for endotherapy.

Patients

This study involved 10 patients (median age 58.5 years [range 27-82 years], 7 men) with GI leaks from perforations, fistulas, and anastomotic dehiscence. Two gastric, 2 duodenal, and 6 colonic leaks were treated with OTSC. The diameter of leaks ranged between 7 and 20 mm.

Interventions

OTSC devices were used to seal the GI leaks. Then Gastrografin (Bayer AG, Germany) was introduced via the endoscope and complete sealing confirmed under fluoroscopy. Patients underwent a second endoscopic examination 3 months later.

Main Outcome Measurement

Complete sealing of the leak.

Results

Complete sealing of leaks was achieved by using OTSC alone in 8 of 10 patients. For one patient, successful endoscopic management was completed by placing two additional covered stents. Only one patient required surgical repair of the leak. Endoscopic examination 3 months after treatment confirmed that leaks in 8 of 9 endoscopically treated patients were healed, and the patients did not have recurrence of the leaks or complications from the OTSC devices. One patient died from neoplastic progression before the second endoscopy could be performed.

Limitations

Uncontrolled study.

Conclusions

The OTSC system appears to be a useful device in the management of larger GI leaks in a variety of clinical scenarios.

El texto completo de este artículo está disponible en PDF.

Abbreviations : NOTES, OTSC


Esquema


© 2010  American Society for Gastrointestinal Endoscopy. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 72 - N° 4

P. 881-886 - octobre 2010 Regresar al número
Artículo precedente Artículo precedente
  • Safety and efficacy of ERCP after recent myocardial infarction or unstable angina
  • Borko Nojkov, Mitchell S. Cappell
| Artículo siguiente Artículo siguiente
  • Life-threatening GI hemorrhage caused by atrioesophageal fistula: a rare complication after catheter ablation for atrial fibrillation
  • Michael J. Baker, Purvi C. Panchal, Patricia A. Allenby

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.