Suscribirse

Component impingement in total hip arthroplasty: Frequency and risk factors. A continuous retrieval analysis series of 416 cup - 09/04/11

Doi : 10.1016/j.otsr.2010.12.004 
E. Marchetti a, b, N. Krantz a, b, C. Berton a, b, D. Bocquet a, b, N. Fouilleron a, b, H. Migaud a, b, , J. Girard a, b
a Lille–Nord-de-France University, 59000 Lille, France 
b Orthopaedic department, Roger-Salengro Hospital, Lille University Hospital, Place de Verdun, 59037 Lille, France 

Corresponding author. Tel.: +33 0320 446 828; fax: +33 0320 446 607.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Summary

Introduction

Impingement is a factor of failure in total hip replacement (THR), causing instability and early wear. Its true frequency is not known; cup-retrieval series reported rates varying from 27 to 84%.

Hypothesis

The hypothesis was that a large continuous series of THR cup removals would help determine the frequency of component impingement.

Objectives

The hypothesis was tested on a continuous retrospective series of cups removed in a single center, with a secondary objective of identifying risk factors.

Materiel and methods

Macroscopic examination looked for component impingement signs in 416 cups retrieved by a single operator between 1989 and 2004. Risk factors were investigated by uni- and multivariate analyses in the 311 cases for which there were complete demographic data. In these 311 cases, removal was for aseptic loosening (131 cases), infection (43 cases), instability (56 cases), osteolysis (28 cases) or unexplained pain (48 cases); impingement was explicitly implicated in only five cases (1.6%), always with hard-on-hard bearing components.

Results

Impingement was found in 214 of the 416 cups (51.4%) and was severe (notch>1mm) in 130 (31.3%). In the subpopulation of 311 cups, impingement was found in 184 cases (59.2%) and was severe in 109 (35%). Neither duration of implant use nor cup diameter or frontal orientation emerged as risk factors. On univariate analysis, impingement was more frequently associated with revision for instability, young patient age at THR, global hip range of motion >200° or use of an extended femoral head flange (or of an elevated antidislocation rim liner), and was more severe in case of head/neck ratio<2. On multivariate analysis, only use of an extended head flange (RR 3.2) and revision for instability (RR 4.2) remained as independent risk factors for impingement.

Discussion

Component impingement is frequently observed in cups after removal, but is rarely found as a direct indication for revision, except in case of hard-on-hard friction couples (polyethylene being the most impingement-tolerant material). Systematic use of extended head flanges and elevated antidislocation rims is not to be recommended, especially in case of excessive ROM. A good head/neck ratio should be sought, notably by increasing the head diameter in less impingement-tolerant hard-on-hard friction couples. Although not identified as a risk factor in the present study, implant orientation should be checked; computer-assisted surgery can be useful in this regard, for adaptation to the patient’s individual range-of-motion cone.

Level of evidence

Level IV, retrospective observational study.

El texto completo de este artículo está disponible en PDF.

Keywords : Impingement, Instability, Wear, Hip prosthesis, Dislocation, Retrieval analysis


Esquema


© 2011  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 97 - N° 2

P. 127-133 - avril 2011 Regresar al número
Artículo precedente Artículo precedente
  • Stem subsidence after total hip revision: 183 cases at 5.9 years follow-up
  • J. Girard, O. Roche, G. Wavreille, F. Canovas, P. Le Béguec
| Artículo siguiente Artículo siguiente
  • Infected total hip arthroplasty treated by an irrigation-debridement/component retention protocol. A prospective study in a 12-case series with minimum 2 years’ follow-up
  • S. Klouche, L. Lhotellier, P. Mamoudy

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.