Suscribirse

Do tantalum components provide adequate primary fixation in all acetabular revisions? - 17/05/10

Doi : 10.1016/j.otsr.2009.11.014 
X. Flecher a, , W. Paprosky b, J.-C. Grillo a, J.-M. Aubaniac a, J.-N. Argenson a
a Osteoarthritis Treatment Surgical Center, South Marseille Teaching Hospital Center Sainte-Marguerite Hospital, 270, boulevard Sainte-Marguerite, BP 29, 13274 Marseille, France 
b Department of Orthopedic Surgery, Central Dupage Hospital, 25N Winfield Road, Winfield, IL 60290, United States of America 

Corresponding author. Tel.: +33 4 91 74 50 12; fax: +33 4 91 74 15 91.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Summary

Introduction

A number of studies have reported favorable results of cementless fixation in acetabular revisions. Nevertheless, the implant bearing on more than 50% of the patient’s bone and good primary stability are required. The objective of this study was to determine whether the use of tantalum implants could provide stable reconstruction for any type of acetabular revision.

Patients and methods

This study investigated 72 hips (71 patients) implanted with tantalum cups, some with augments, and morselized graft material. The mean age was 60 years (range, 34–84 years). There were 30 males and 41 females. The mean weight was 71 kg (range, 52–102 kg), the mean height was 1.68 m (range, 1.52–1.84 m). Twenty-five revisions were bipolar. Six revisions were performed for infected acetabular loosening. The mean follow-up was 4 years (range, 2–6 years).

Results

The mean Merle d’Aubigné score at follow-up was 15.8 points (range, 9–18 points). According to the Paprosky classification of acetabular bone defects, there were 13 type 1 (18%), 14 type 2A (19.5%), 14 type 2B (19.4%), 23 type 3A (31.9%), and eight type 3B (11.2%) acetabular defects, four of which had pelvic discontinuity. Postoperatively, the position of the hip’s center of rotation in relation to Köhler’s teardrop was 22 mm (range, 5–41 mm) vertically (normal, < 25 mm) and 39mm (range, 13–55 mm) horizontally (normal, < 35 mm). The mean acetabular inclination was 40° (range, 20°–63°). The radiographic analysis found no radiolucent line after 1 year and up to the last follow-up. None of the patients required revision for acetabular loosening. Three hips were revised for instability. Two retentive liners and a dual-mobility cup were cemented in the cups that were left in place.

Discussion and conclusion

Given their mechanical properties (coefficient of friction, porosity), tantalum implants provide a stable primary cementless fixation without compromising the center of rotation and without necessarily requiring a structural graft. A single implant range can therefore be used for any type and severity of bone loss and for all types of acetabular reconstruction. Longer follow-up is nevertheless necessary to confirm these encouraging results.

Level of evidence

Level IV, historical series.

El texto completo de este artículo está disponible en PDF.

Keywords : Total hip arthroplasty, Revision, Acetabular loosening, Cementless, Biological fixation, Tantalum, Trabecular metal


Esquema


© 2010  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 96 - N° 3

P. 235-241 - mai 2010 Regresar al número
Artículo precedente Artículo precedente
  • Garden I femoral neck fractures in patients 65 years old and older: Is conservative functional treatment a viable option?
  • J.-M. Buord, X. Flecher, S. Parratte, L. Boyer, J.-M. Aubaniac, J.-N. Argenson
| Artículo siguiente Artículo siguiente
  • Comparative study of intraoperative knee flexion with three different TKR designs
  • A. Wajsfisz, D. Biau, P. Boisrenoult, P. Beaufils

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.