Optimisation de la date limite d’utilisation des dispositifs médicaux réutilisables. Expérience dans un établissement de santé - 22/02/24

Optimizing the use-by date of reusable medical devices. Experience in a healthcare establishment

Doi : 10.1016/j.pharma.2023.12.007 
Manon Vidor , Christelle Raguenes, Hélène Astier, Gilles Gbaguidi, Sarah Le Tohic
 Service de pharmacie hospitalière, hôpital d’instruction des Armées Laveran, 34, boulevard Laveran, 13013 Marseille, France 

Auteur correspondant.

Highlights

Le stockage, étape la plus à risque de la perte de stérilité dans le temps.
Sachets : passage de la date limite d’utilisation de deux à quatre mois.
Pliages : passage de la date limite d’utilisation de deux à six mois.
Conteneurs : maintien de la date limite d’utilisation à six mois.
Impact positif de l’augmentation de la date limite d’utilisation.

El texto completo de este artículo está disponible en PDF.

Résumé

Objectifs

Les dispositifs médicaux réutilisables (DMR) stérilisés ont une date limite d’utilisation (DLU), au-delà de laquelle sa stérilité n’est plus garantie. Sa détermination ne répond actuellement à aucun consensus. L’objectif est de réévaluer la DLU des DMR, grâce à une analyse de risque et à une évaluation du maintien de l’état stérile des DMR dans le temps.

Méthodes

L’analyse de risque portait sur les étapes, dont le dysfonctionnement peut compromettre la stérilité des DMR dans le temps : le conditionnement, le transport et le stockage. Une cartographie des risques est réalisée suivant la méthodologie préconisée par la Haute Autorité de santé. En s’appuyant sur la norme NF EN ISO 11737, l’évaluation du maintien de l’état stérile était contrôlée sur des DMR après deux, quatre et six mois de stockage et sur des DMR périmés depuis plus d’un an.

Résultats

L’analyse de risque a identifié quatre défaillances et soixante-huit causes potentielles. L’étape jugée la plus sensible est le stockage car elle regroupe la plupart des causes critiques et majeures. Des actions d’amélioration ont été proposées, comme la définition d’un plan de maintenance des conteneurs. En parallèle, 256 DMR ont été testés. Les cultures sont restées stériles pour tous les conteneurs, pour les pliages testés à 6 mois et plus et pour les sachets testés à 2 et 4 mois et à plus d’un an de stockage.

Conclusions

La DLU a été allongée à 4 mois pour les sachets, à 6 mois pour les pliages et elle a été maintenue à six mois pour les conteneurs.

El texto completo de este artículo está disponible en PDF.

Summary

Objectives

Sterilized reusable medical devices have a use-by date, after which sterility is no longer guaranteed. There is currently no consensus on how this should be determined. The aim is to re-evaluate the expiry date of reusable medical devices, by means of a risk analysis and an assessment of the maintenance of the sterile state of reusable medical devices over time.

Methods

The risk analysis focused on the stages whose malfunction could compromise the sterility of reusable medical devices over time: packaging, transport and storage. Risk mapping was carried out in accordance with the methodology recommended by the French Health Authority. Based on standard NF EN ISO 11737, the assessment of the maintenance of the sterile state was checked on reusable medical devices after two, four and six months storage and on reusable medical devices that had expired more than a year previously.

Results

The risk analysis identified four failures and sixty-eight potential causes. The most sensitive stage was storage, which accounted for most of the critical and major causes. Improvement actions were proposed, such as the definition of a container maintenance plan. At the same time, 256 reusable medical devices were tested. The cultures remained sterile for all the containers, for folded products tested at 6 months and more and for the sachets tested at 2 and 4 months and at more than one year of storage.

Conclusions

The DLU has been extended to 4 months for sachets, 6 months for folded products and maintained at six months for containers.

El texto completo de este artículo está disponible en PDF.

Mots clés : Dispositif médical, Analyse des risques et maîtrise des points critiques, Évaluation de processus

Keywords : Medical device, Hazard analysis and critical control point, Process evaluation


Esquema


© 2023  Académie Nationale de Pharmacie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 82 - N° 2

P. 292-305 - mars 2024 Regresar al número
Artículo precedente Artículo precedente
  • Optimizing pharmaceutical management of clinical trials
  • Akram Farhat, Aline Voidey, Isabelle Sommer, Laurent Carrez, Farshid Sadeghipour
| Artículo siguiente Artículo siguiente
  • Étude comparative des processus de production et de contrôle analytique des préparations magistrales de mélanges nutritifs parentéraux en néonatalogie
  • Alizée Bosson, Marjorie Durand, Roseline Mazet, Mireille Jouannet-Romaszko, Philip Chennell, Freddy Mounsef, Thierry Quessada, Luc Choisnard, Pierrick Bedouch

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2025 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.