Experimental study of risk of medial hinge fracture during distal femoral varus osteotomy - 20/05/23

Doi : 10.1016/j.otsr.2022.103527 
Thomas Mereb a, b, Henri Favreau a, Matthieu Ollivier c, Hamdi Jmal b, François Bonnomet a, Nadia Bahlouli b, Pierre Martz d, e, Matthieu Ehlinger a, b,
a Service de Chirurgie Orthopédique et de Traumatologie, Hôpital de Hautepierre, Hôpitaux Universitaires de Strasbourg, 1, avenue Molière, 67098 Strasbourg cedex, France 
b Laboratoire ICube, CNRS UMR 7357, 2-4, rue Boussingault, 67000 Strasbourg, France 
c Département de Chirurgie Orthopédique et Traumatologie, Aix-Marseille Université, AP–HM, CNRS, ISM, Hôpital Sainte-Marguerite, Institut de la Locomotion, Marseille, France 
d CHU Dijon, Service d’Orthopédie Traumatologie, 14, rue Paul-Gaffarel, 21000 Dijon, France 
e Unité Inserm UMR CAPS U1093, UFR STAPS, Campus Universitaire, BP 27877, 21078 Dijon cedex, France 

Corresponding author at: Service de Chirurgie Orthopédique et de Traumatologie, Hôpital de Hautepierre, Hôpitaux Universitaires de Strasbourg, 1, avenue Molière, 67098 Strasbourg cedex, France.Service de Chirurgie Orthopédique et de Traumatologie, Hôpital de Hautepierre, Hôpitaux Universitaires de Strasbourg1, avenue MolièreStrasbourg cedex67098France

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Introduction

Lateral opening wedge distal femoral osteotomy (LOWDFO) is indicated for isolated lateral osteoarthritis in the valgus morphotype. Medial hinge fracture is a factor for poor prognosis. The present study had two aims: (1) to assess the impact of a temporary K-wire on hinge fracture risk; and (2) to assess the impact of LOWDFO opening speed.

Hypothesis

The main study hypothesis was that a temporary hinge K-wire reduces hinge fracture risk. The second hypothesis was that faster opening speed increases fracture risk.

Material and method

Twenty femurs were produced by 3D printing from a CT database, reproducing LOWDFO anatomy. The ABS® polymer showed the same breaking-point behavior as human bone. Ten specimens were included in the “K-wire” group (KW+) and 10 in the “No K-wire” group (KW−). To determine high and low speed, a motion-capture glove was used by 2 operators, providing 3D modeling of the surgeon's hand. High speed was defined as 152mm/min and low speed as 38mm/min. The KW+ and KW− groups were subdivided into high- and low-speed subgroups (HS, LS) of 5 each. Compression tests were conducted using an Instron® mechanical test machine up to hinge fracture. The main endpoint was maximum breaking-point force (N); the secondary endpoints were maximum displacement (mm) and maximum speed (min) at breaking point.

Results

The K-wire significantly increased maximum breaking-point force (LS, 143.08N vs. 93.71N, p<0.01; and HS, 186.98N vs. 95.22N, p<0.01), but not maximum displacement (LS, 26.17mm vs. 24.11mm, p=0.31; and HS 26.18mm vs. 23.66mm, p=0.14) or maximum time (LS, 27.07s vs. 24.94s, p=0.31; and HS, 5.24s vs. 4.73s, p=0.14). Speed did not affect maximum force (KW+, 143.08N vs. 186.98N, p=0.06; and KW−, 93.71N vs. 95.22N, p=0.42) or maximum displacement (KW+, 26.17mm vs. 26.18mm, p=1; and KW−, 24.11mm vs. 23.66mm, p=0.69). Only maximum time was greater at low speed (KW+, 27.07s vs. 5.24s, p>0.01; and KW−, 24.94s vs. 4.73s, p<0.01), which is obvious for constant distance.

Discussion

The first study hypothesis was confirmed, with significantly lower hinge fracture risk with the K-wire, independently of opening speed. The second hypothesis was not confirmed. The study was performed under strict experimental conditions, unprecedented to our knowledge in the literature. However, complementary clinical studies are needed to confirm the present findings.

Level of evidence

IV, experimental study.

El texto completo de este artículo está disponible en PDF.

Keywords : Distal femoral osteotomy, Hinge fracture, Maximum force, Biomechanical study


Esquema


© 2022  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 109 - N° 4

Artículo 103527- juin 2023 Regresar al número
Artículo precedente Artículo precedente
  • Biplanar high tibial osteotomy for the combined correction of varus and posterior tibial slope malalignment
  • Sebastian Müller, Karl-Heinz Frosch, Jannik Frings, Markus Berninger, Matthias Krause
| Artículo siguiente Artículo siguiente
  • Risk of simultaneous and delayed breakage of total hip replacement ceramic implants in patients with trauma induced periprosthetic fracture and acetabular shell loosening
  • Robin Lévêque, Laurent Sedel, Rémy Nizard, Pascal Bizot, Guillaume-Anthony Odri

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2025 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.