Doi : 10.1016/j.otsr.2021.103136 
Antonio Klasan a, b, c, , Ali Bayan a, Ian Holdaway d, William J. Farrington a
a North Shore Hospital, 124, Shakespeare Road, Takapuna, 0620 Auckland, New Zealand 
b Kepler University Hospital Linz, Krankenhausstr. 9, 4020 Linz, Austria 
c Johannes-Kepler University Linz, Altenbergerstr. 69, 4040 Linz, Austria 
d Auckland City Hospital, 2, Park Road, Grafton, 1023 Auckland, New Zealand 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

Three-dimensional printing of implants allows the ability to produce implants and interfaces which theoretically better mimic “normal” bone behaviour, leading to a possible reduction in stress shielding thus maintaining bone mineral density (BMD). This issue was not investigated in vivo using bone scan and different bearings; therefore, we did a prospective study aiming to answer: 1) is there a loss of BMD around the 3D printed trabecular titanium cup, when compared to the native hip?; 2) does liner type influence the BMD changes around the acetabulum when a 3D printed trabecular titanium cup is used?

Hypothesis

BMD changes around the acetabulum are not influenced by the liner type, and the cup will be associated with a reduction in BMD when compared to the native hip.

Material and methods

This is a prospective observational study of patients receiving a primary total hip arthroplasty. A 3D printed trabecular titanium uncemented acetabular component was used in all cases. All patients received a ceramic femoral head, with either a ceramic or polyethylene acetabular liner. BMD measurements using DXA were performed at 6 weeks, 6, 12 and 24 months after surgery to evaluate remodeling changes. The 3 acetabular regions of interest (ROI) of DeLee and Charnley were used for serial comparisons of peri-acetabular BMD. The study was powered as a non-inferiority study with the principle variables compared using a two-step repeated analysis of variance.

Results

A total of 48 consecutive patients were included in the study, with all patients completing their 2 year follow-up. There were no failures, revisions or complications within this cohort. We found no statistically significant difference in the BMD change scores between the operated and the native hip in any of the 3 ROI zones. We found no differences in BMD scores when comparing ceramic to polyethylene acetabular liners, head sizes and BMI.

Discussion

This study shows a similar pattern of BMD behaviour around a 3D printed cup when compared to the contralateral native hip. We were unable to show a clinical or radiological difference between the bearing material, head size, or BMI when used with this type of acetabular component.

Level of evidence

III; prospective comparative study.

El texto completo de este artículo está disponible en PDF.

Keywords : Total hip arthroplasty, Cementless, 3D printing, DXA, Bone density


Esquema


© 2021  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 109 - N° 1

Artículo 103136- février 2023 Regresar al número
Artículo precedente Artículo precedente
  • Wear rate and osteolysis in two types of second-generation annealed highly cross-linked polyethylene in total hip arthroplasty: A retrospective comparative study with a minimum of five years
  • Naoki Yamamoto, Ryohei Takada, Tetsuya Jinno, Kazumasa Miyatake, Naoto Watanabe, Hideyuki Koga, Toshitaka Yoshii, Kazuyoshi Yagishita, Atsushi Okawa
| Artículo siguiente Artículo siguiente
  • Doctor, what does my ceramic-on-ceramic hip arthroplasty weigh?
  • Simon Marmor, Emmanuel De Panafieu, Antoine Mouton, Wilfrid Graff, Vincent Le Strat, Luc Lhotellier, Younes Kerroumi, Emmanuel Gibon

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2025 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.