Comparison of peak skin dose and dose map obtained with real-time software and radiochromic films in patients undergoing abdominopelvic embolization - 02/07/22

Doi : 10.1016/j.diii.2022.01.010 
Joël Greffier a, b, , Asmaa Belaouni a, Djamel Dabli a, b, Jean Goupil a, Romain Perolat a, Philippe Akessoul a, Tarek Kammoun a, Adel Hoballah a, Jean Paul Beregi a, Julien Frandon a
a Department of Medical Imaging, CHU Nimes, Univ Montpellier, Medical Imaging Group Nimes, EA 2992, 30000 Nîmes, France 
b Department of Medical Physics, CHU Nimes, Univ Montpellier, 34000 Montpellier, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Highlights

A substantial correlation is found between the peak skin dose computed with the Dose Tracking System software and that obtained using radiochromic film.
Similar dose maps are found with the Dose Tracking System software and radiochromic films for patients undergoing abdominal embolization.
For all patients undergoing abdominal embolization, peak skin doses are significantly lower with Dose Tracking System software than with radiochromic film.

El texto completo de este artículo está disponible en PDF.

ABSTRACT

Purpose

The purpose of this study was to compare peak skin dose (PSD) and dose map calculated by Dose-Tracking-System® (DTS) software and measured with radiochromic films in patients undergoing abdominopelvic embolization.

Material and methods

The PSD measured by radiochromic films (PSDFilm) or calculated by DTS software (PSDDTS) were compared in patients who underwent abdominopelvic embolization between September 2020 and April 2021. Concordance between PSDFilm and PSDDTS was computed using the Lin's concordance correlation coefficient and the clinical concordance using Bland Altman analysis. PSD values were compared using the paired Mann-Whitney-Wilcoxon test.

Results

A total of 40 patients were included. There were 32 men and 8 women with a mean age of 73.0 ± 14.6 (SD) years (age range: 30–92 years). Median PSDFilm was 756 mGy (IQR: 390; 1094) and median PSDDTS was 768 mGy (IQR: 421; 1076), resulting in a median difference of -5% (IQR: -10%; 0%) between PSDFilm and PSDDTS (P = 0.024). The concordance correlation between PSDFilm and PSDDTS was substantial in patients (0.986; 95% CI: 0.977–0.992). Bland Altman analysis showed that PSDDTS was underestimated compared to PSDFilm by -36 mGy (95% CI: -68–-4). Visually similar dose maps were found with DTS and radiochromic films.

Conclusion

Real-time DTS software allows computing PSD with high accuracy and generating adequate dose map. PSDDTS was slightly underestimated compared to PSDFilm requiring the use of a correction factor for the PSDDTS to avoid lack of follow-up for some patients.

El texto completo de este artículo está disponible en PDF.

Keywords : Embolization, Fluoroscopy, Film dosimetry, Interventional radiology, Radiochromic films

Abbreviations : CI, CT, DLP, DTS, IRP, IQR, Ka, KAP, PACS, PSD, Q1, Q3, SD, SDC, TACE, VERIDIC


Esquema


© 2022  Société française de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 103 - N° 7-8

P. 338-344 - juillet 2022 Regresar al número
Artículo precedente Artículo precedente
  • Retrospective analysis of dose delivered to the uterus during CT examination in pregnant women
  • Djamel Dabli, Mélinée Linard, Quentin Durand, Julien Frandon, Fabien de Oliveira, Jean Paul Beregi, Joël Greffier
| Artículo siguiente Artículo siguiente
  • Left ventricular diastolic early and late filling quantified from 4D flow magnetic resonance imaging
  • Yousef Alattar, Gilles Soulat, Umit Gencer, Emmanuel Messas, Emilie Bollache, Nadjia Kachenoura, Elie Mousseaux

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2025 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.