Suscribirse

Screw misplacement in percutaneous posterior pelvic iliosacral screwing with and without navigation: A prospective clinical study of 174 screws in 127 patients - 13/04/22

Doi : 10.1016/j.otsr.2022.103213 
Mehdi Boudissa a, b, , Delphine Carmagnac a, Gaël Kerschbaumer a, Sébastien Ruatti a, Jérôme Tonetti a, b
a Service de chirurgie orthopédique et traumatologique, centre hospitalier universitaire de Grenoble, université Grenoble Alpes, 38700 La Tronche, France 
b TIMC-IMAG lab, Université. Grenoble Alpes, CNRS UMR 5525, 38700 La Tronche, France 

Corresponding author at: Service de chirurgie orthopédique et traumatologique, CHU Grenoble, hôpital Nord, université Grenoble Alpes, boulevard de la Chantourne, 38700 La Tronche, France.Service de chirurgie orthopédique et traumatologique, CHU Grenoble, hôpital Nord, université Grenoble Alpesboulevard de la ChantourneLa Tronche38700France

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

Recent studies of iliosacral screw fixation performed using intraoperative navigation systems have shown promising results. The Surgivisio Platform is a new-generation three-dimensional intraoperative navigation tool that has been used at our institution for 2years. The aim of this prospective study was to assess the contribution of navigation in terms of iliosacral screw positioning accuracy and of radiation exposure, by comparing outcomes with vs. without navigation.

Hypothesis

Navigation allows more accurate percutaneous iliosacral screw positioning regardless of the type of screw fixation (with sacral dysmorphism and/or cemented screw fixation and/or multiple screw fixation).

Materials and methods

Between January 2018 and December 2019, consecutive patients who underwent percutaneous iliosacral screw fixation of pelvic ring fractures without vertical instability were included in this single-centre prospective study. Screw position accuracy was evaluated by postoperative high-resolution computed tomography (HRCT). Operative time, radiation dose, and complications were recorded.

Results

We included 127 patients with 174 iliosacral screws, of which 129 were positioned under fluoroscopic guidance and 45 using navigation. According to the modified Gras classification, 7% (12/174) of the screws were incorrectly positioned and 2% (4/174) required repositioning. The frequency of screw malposition was not significantly different between the fluoroscopy and navigation groups (8.5%, 11/129 vs. 2.2%, 1/45, respectively; p=0.19). However, screw position in dysmorphic sacra was significantly better with navigation (p=0.04), whereas no significant difference in final screw position was found for cemented or multiple screw implantations. In the navigation group, the operative time was significantly longer (28.2min vs. 21.6min, p=0.003), and the mean dose-area product significantly greater (6.6Gy·cm2 vs. 4.9Gy·cm2, p=0.02). The complication rates were not different between the two groups.

Conclusion

In patients who have pelvic ring fractures without vertical instability, navigation of percutaneous iliosacral screw placement using the Surgivisio Platform improves screw positioning in dysmorphic sacra, at the cost of a longer operative time and greater radiation exposure of the patient.

Level of evidence

II, prospective study.

El texto completo de este artículo está disponible en PDF.

Keywords : Iliosacral screw fixation, Computer-assisted surgery, Radiation exposure, Sacral dysmorphism, Surgivisio®


Esquema


© 2022  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 108 - N° 2

Artículo 103213- avril 2022 Regresar al número
Artículo precedente Artículo precedente
  • Long-term survival of the Lefèvre retentive cup: 12-year follow-up analysis of 466 consecutive cases
  • Hoel Letissier, Aurélien Barbier, Ludovic Tristan, Frédéric Dubrana, Christian Lefèvre, Arnaud Clavé
| Artículo siguiente Artículo siguiente
  • Iliosacral screw fixation of pelvic ring disruption with tridimensional patient-specific template guidance
  • Feng Liu, Junyi Yu, Hongqi Yang, Lihong Cai, Li Chen, Qing Lei, Pengfei Lei

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.