Suscribirse

Ceramic-on-ceramic total hip arthroplasty is superior to metal-on-conventional polyethylene at 20-year follow-up: A randomised clinical trial - 30/01/21

Doi : 10.1016/j.otsr.2020.102744 
Pascal-André Vendittoli a, b, c, , Maged Shahin a, Charles Rivière b, d, e, Janie Barry a, Pauline Lavoie c, Nicolas Duval c
a Hôpital Maisonneuve-Rosemont, Université de Montréal, Département de chirurgie, 5415 boulevard de l’Assomption, Montréal, Québec H1T 2M4, Canada 
b Personalized Arthroplasty Society, Montreal, Canada 
c Clinique Orthopédique Duval, Laval, Québec, Canada 
d MSK Lab – Imperial College de Londres. Sir Michael Uren Hub, White City Campus, 86 Wood Ln, London W12 0BZ, UK 
e Centre de l’Arthrose, Clinique du Sport, 04–06, rue Georges-Negrevergne, 33700 Mérignac, France 

Corresponding author at: Université de Montréal, Hôpital Maisonneuve-Rosemont, 5415 boulevard L’Assomption, Montréal, Québec H1T 2M4, Canada.Université de Montréal, Hôpital Maisonneuve-Rosemont5415 boulevard L’Assomption, MontréalQuébec H1T 2M4Canada

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

Metal-on-conventional polyethylene (MoPc) bearing wear-related biological reactions in total hip arthroplasty (THA) continue to raise concerns among young, active patients. Ceramic-on-ceramic (CoC) bearings may offer improved outcomes in this patient population.

Questions/purposes

The aim of this study was to determine if, more than 20years postoperatively, there is a difference between MoPc and CoC THA in terms of (1) survivorship, (2) related complications, (3) radiographic signs of wear, and (4) functional scores.

Hypothesis

CoC bearing THAs have superior clinical results compared to MoPc THAs.

Patients and methods

A total of 140 hips in 116 patients with a mean age of 42years were randomised to receive CoC or MoPc THA between 1996 and 2001. Sixty-nine hips in 58 patients received MoP and 71 hips in 68 patients received CoC. Revision rate, WOMAC score, and radiological signs of osteolysis and loosening were compared at last follow-up.

Results

After a mean follow-up of 21years (19–23), 40 patients (48 THAs; 34%) had died and 6 patients (6 THAs; 4%) were lost to follow-up. Aseptic revision rate was significantly higher in the MoPc group (17/69; 24.6%) versus CoC (2/71; 2.8%; p<0.001). Kaplan–Meier survivorship estimator with revision for aseptic reasons was 73.6% (95% CI: 63.3–84.9%) for MoPc and 96.9% (95% CI: 92.8–100%) for CoC (p<0.001). On radiographic evaluation, 13% (3/23) MoPc were considered loose versus no CoC, and 61% (14/23) MoPc versus 6% (2/33) CoC showed osteolytic signs (p<0.001). CoC had better mean WOMAC scores than MoPc (11.0 vs. 19.4; p=0.048). No ceramic fracture was observed.

Conclusion

In this RCT, CoC bearings provided excellent results and were safer than MoPc bearings at more than 20-year follow-up. The long-term in vivo behaviour of CoC bearing makes it a great THA option for middle-aged patients and should be compared to newer polyethylene bearings.

Level of evidence

I.

El texto completo de este artículo está disponible en PDF.

Keywords : Total hip arthroplasty, Ceramic-on-ceramic, Survival, Revision, Complications


Esquema


© 2020  The Authors. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 107 - N° 1

Artículo 102744- février 2021 Regresar al número
Artículo precedente Artículo precedente
  • Does load-bearing materials influence hip capsule thickness in total hip replacement? An MRI case-matched study
  • Guillaume-Anthony Odri, Julien Sanchez, Jean-Michel Sverzut, Jean-Denis Laredo, Laurent Sedel, Pascal Bizot
| Artículo siguiente Artículo siguiente
  • Histochemical identification of wear debris released by alumina-on-alumina hip prostheses in the periprosthetic tissues
  • Louis Rony, Florence Pascaretti-Grizon, Laurent Hubert, Daniel Chappard

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2025 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.