Optimal configuration for stability and magnetic resonance imaging quality in temporary external fixation of tibial plateau fractures - 12/11/20

Doi : 10.1016/j.otsr.2019.12.025 
Massimo Max Morandi a, Alberto Simoncini b, Claire Hays a, John Garrett b, R. Shane Barton a, Andreas Chen a, Giovanni F. Solitro a,
a Department of Orthopaedic Surgery, Louisiana State University Health Science Center, Shreveport, LA, United States 
b Department of Radiology, Louisiana State University Health Science Center, Shreveport, LA, United States 

Corresponding author. Department of Orthopaedic Surgery, Louisiana State University Health Science Center, Shreveport, LA, United States.Shreveport, LAUnited States

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Highlights

Tibial plateau fractures may require both external fixation and MRI so we tested constructs to find the most suitable design in both aspects.
Image artifact from external fixators is thought to affect diagnostic quality of MRI scans but we found no significant diagnostic interference.
External fixator stability improves fracture healing and varies with construct so we compared constructs to find the most stable design.

El texto completo de este artículo está disponible en PDF.

Abstract

Introduction

Temporary external fixation has been widely utilized in the stabilization of plateau fractures while waiting for an optimization of the soft tissue conditions before subsequent permanent internal fixation. Simultaneously, MRI is beneficial in the assessment of concomitant damage to ligaments and menisci so that these injuries could be promptly identified, and surgical planning executed at the time of definitive fixation of the bony injury. Increasing numbers of side-bars and pins have been previously suggested to increase frame rigidity, but at the same time, several studies have indicated the presence of MRI artifacts which may obscure key anatomical structures, even when MRI-compatible fixation devices are used. This study aims to identify, among six potential configurations, the construct that maximizes stability while most minimizing the number of MRI artifacts generated among different configurations commonly used.

Hypothesis

There is one or more configurations among the others that maximize stability while preserving a clinically acceptable level of MRI quality.

Material and methods

Six constructs were recreated on cadaveric specimens and identified by the disposition of the bars: H, Anterior, Flash, Hashtag, Rhomboid, and Diamond. Stage one evaluated the amount of artifact produced during MRI on instrumented cadaveric legs, as well as the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) at five specific regions of interest. Stage two assessed the amount of compressional and torsional stiffness of the configurations on bone surrogate models.

Results

Image artifacts were not detected within the knee joint for all considered constructs. In terms of SNR The H, Anterior, Hashtag, and Diamond configurations were not significantly different from their control (p>0.366) while the others were significantly different (p<0.03). The values of CNR found for the H and Hashtag configurations were not significantly different from their controls (p>0.07) while the remaining configurations were significantly different (p<0.03). In compression, the H and Diamond configurations had similar stiffness (p=0.468) of 35.78N/mm and 31.44N/mm, respectively, and were stiffer than the other configurations. In torsion, the constructs have shown different stiffness (p<0.001) with a minimum value of 0.66 Nm/deg for the Rhomboid configuration, which was significantly less stiff than the Anterior configuration (1.20 Nm/deg [p<0.001]). There was no difference between the Diamond and H configurations (p=0.177) or between them and the Hashtag configuration (p=0.215).

Discussion

An external fixator construct directly bridging the femur and tibia without interconnections is the most stable and produces MRI scans without image artifacts that would interfere with diagnostic quality.

Level of evidence

V, basic science study, diagnostic imaging and mechanical testing.

El texto completo de este artículo está disponible en PDF.

Keywords : Tibial plateau fracture, External fixation stability, Magnetic resonance imaging, Image artifact


Esquema


© 2020  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 106 - N° 7

P. 1405-1412 - novembre 2020 Regresar al número
Artículo precedente Artículo precedente
  • Early mortality and morbidity of odontoid fractures after 70 years of age
  • Jeremy Allia, Hugo Darmanté, Laurent Barresi, Fernand De Peretti, Christophe Trojani, Nicolas Bronsard
| Artículo siguiente Artículo siguiente
  • Contribution of the remaining attachment index in the management of Vancouver B1 periprosthetic hip fracture
  • Tsiry Andriamananaivo, Guillaume-Anthony Odri, Matthieu Ollivier, Lucas Mattesi, Arthur Renault, Frédéric Rongieras, Sébastien Pesenti, Mathieu Severyns

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2025 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.