Abstract

Background

Rotating-hinge knee replacements are usually reserved for revision surgeries, when the extent of soft tissue loss makes a constrained implant more suitable. They remain an uncommon choice in primary surgery when the soft tissue loss is not as extensive.

Methods

We completed a systematic review and meta-analysis to assess patients who underwent a Total Knee Replacement (TKR) with the rotating-hinge prosthesis in the primary setting. We searched PubMed and Embase for articles published in the ten years prior June 2017: Prosthesis survival rates, causes of failure, and clinical/functional scores were the primary outcomes. Twenty-one articles met the inclusion criteria for meta-analysis. Articles were grouped into (1) non-tumour (n=11) and (2) tumour indications (n=10). Survival data was summarized in forest plots, generated using Stata.

Results

We found that for certain indications the prosthesis has impressive survival rates and functional outcomes. Short-term (1–5 year) prosthesis survival in non-tumour cases was 92% (95% CI, 87–98%) and 77% (95% CI, 68–87%) in tumour cases. Mid-term (6–10 year) survival was 82% (95% CI, 74–89%) and 69% (95% CI, 57–81%) in non-tumour and tumour studies respectively. In analysis of clinical scores, patients showed a significant improvement in their pain score. Infection was the most commonly cited cause of prosthesis failure in both non-tumour and tumour studies, attributing to 31.5% and 37.6% of failures respectively. Aseptic loosening, dislocation and fracture were also commonly cited complications.

Conclusion

We concluded that the rotating-hinge knee prosthesis is a viable option in primary surgery when there is extensive soft tissue destruction surrounding the joint.

Level of evidence

I.

El texto completo de este artículo está disponible en PDF.

Keywords : Rotating hinge TKA, Total knee replacement, Primary surgery, Survival rate, Causes of failure


Esquema


© 2019  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 105 - N° 7

P. 1351-1359 - novembre 2019 Regresar al número
Artículo precedente Artículo precedente
  • Variability in long-term pain and function trajectories after total knee replacement: A cohort study
  • Vikki Wylde, Chris Penfold, Alice Rose, Ashley W. Blom
| Artículo siguiente Artículo siguiente
  • Compared outcomes 16 and 25 years after lateral wedge augmentation trochleoplasty: Rate of recurrent dislocation and progression to osteoarthritis
  • Eloïse Bauduin, Sophie Putman, Henri Migaud, Franck Remy, Emmanuel Debuyzer, Gilles Pasquier

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2025 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.