Are commercially-available precontoured anatomical clavicle plating systems offering the purported superior optimum fitting to the clavicle? A cadaveric analysis and review of literature - 27/09/18

Doi : 10.1016/j.otsr.2018.01.013 
D.E. Bauer a, , A. Hingsammer a, P. Schenk a, L. Vlachopoulos b, M.A. Imam a, P. Fürnstahl b, D.C. Meyer a
a Department of Orthopaedics, Balgrist University Hospital, University of Zurich, 8008 Zurich, Switzerland 
b Computer Assisted Research and Development Group, Balgrist University Hospital, 8008 Zurich, Switzerland 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose

The indication for operative treatment of displaced midshaft clavicle fractures remains controversial. However, if plate fixation is considered, implant prominence and skin irritation are the most common causes for re-operation. Low profile implants as well as closely contouring plates to the individual anatomy may reduce these complications. The aim of this study was to compare the fitting accuracy and implant prominence of 3.5mm pelvic reconstruction plates (PRP) with pre-contoured anatomical clavicle plates (PACP) for midshaft clavicle fractures.

Methods

Three-dimensional data of the largest, median and smallest male and female clavicle of an existing database of 89 cadaveric clavicles were included for analysis. A three-dimensional model of a commercially available PACP was used for digitally positioning of the plate on the segmented clavicles. Three-dimensional printouts of each clavicle were produced and the 3.5mm reconstruction plates were manually bent and positioned by the senior author. Computed tomography scans and three-dimensional reconstructions were then obtained to digitally compare the fitting accuracy and implant prominence.

Results

Pelvic reconstruction plates offered superior fitting accuracy and lower implant prominence compared to PACP. The largest difference in implant prominence was observed in large sized female clavicles and measured 3.6mm.

Conclusion

Both, the less costly PRP plates and commercially available PACP for midshaft fractures of the clavicle demonstrated a clinically acceptable fitting accuracy. The manually bent pelvic-reconstruction plates demonstrated reduced implant prominence with superior fitting. Hypothetically this might contribute to a reduced rate of reoperation.

Level of evidence

Level IV cadaveric study.

El texto completo de este artículo está disponible en PDF.

Keywords : Clavicle, Fracture, Midshaft, Reconstruction plate, Three-dimensional


Esquema


© 2018  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 104 - N° 6

P. 755-758 - octobre 2018 Regresar al número
Artículo precedente Artículo precedente
  • The use of the LARS system in the treatment of AC joint instability – Long-term results after a mean of 7.4 years
  • T.M. Tiefenboeck, S. Boesmueller, D. Popp, S. Payr, J. Joestl, H. Binder, M. Schurz, M. Komjati, C. Fialka, R.C. Ostermann
| Artículo siguiente Artículo siguiente
  • Is reverse total shoulder arthroplasty more effective than hemiarthroplasty for treating displaced proximal humerus fractures in older adults? A systematic review and meta-analysis
  • David Gallinet, Xavier Ohl, Lauryl Decroocq, Choukry Dib, Philippe Valenti, Pascal Boileau, the French Society for Orthopaedic Surgery (SOFCOT) e

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.