Anterior pre-tensioned external fixator for pelvic fractures and dislocations. Initial clinical series - 01/12/16
, M. Lombardo-Torre ⁎
, A. Leiva-Gea
, F.B. Delgado-Rufino
, F. Luna-González 
Abstract |
In the treatment of unstable pelvic ring fractures, external fixators have the limitation of not adequately stabilizing the injured posterior elements. This article presents a novel and simple technique of temporary external fixation of the pelvic ring, able to produce compression of both the anterior and posterior pelvic elements. A curved flexible carbon-fiber rod is used, pre-tensioned before attachment to supra-acetabular Schanz screws. Although more extensive clinical experience is required, favorable preliminary results in a series of 13 patients with unstable pelvic fracture were encouraging: the aim of closing the posterior and anterior elements of the pelvic ring was achieved in all cases treated with this technique, and 12 patients survived. Radiological results were excellent in 3 cases and good in 9 cases. No major complications, such as secondary displacement, vertical re-displacement or deep infection, were observed. Mean operative time was 25min, compatible with emergency management.
El texto completo de este artículo está disponible en PDF.Keywords : Pelvis fracture, Urgent stabilization, Pelvic compression, External fixation, Pre-tensed bar