Évaluation des reproductibilités inter- et intra-observateurs de la mesure du futur foie restant par tomoscintigraphie hépato-biliaire à la 99mTc-mébrofénine avant hépatectomie majeure - 28/03/15

Doi : 10.1016/j.mednuc.2015.01.005 
G. Nivaggioni a , C. Baillet a, A. Béron a, S. Truant b, A. Duhamel c, d, F.R. Pruvot b, d, D. Huglo a, , d
a Service de médecine nucléaire, hôpital Huriez, CHRU de Lille, rue Michel-Polonovski, 59037 Lille cedex, France 
b Service de chirurgie digestive et transplantation, hôpital Huriez, CHRU de Lille, rue Michel-Polonovski, 59037 Lille cedex, France 
c Plateforme d’aide méthodologique, CHRU de Lille, 2, avenue Oscar-Lambret, 59037 Lille cedex, France 
d Université de Lille, faculté de médecine, 1, place de Verdun, 59000 Lille, France 

Auteur correspondant.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

pages 10
Iconographies 7
Vidéos 0
Autres 0

Résumé

Introduction

La scintigraphie hépato-biliaire à la mébrofénine avant hépatectomie majeure est actuellement étudiée pour la détection des patients à risque d’insuffisance hépatique, généralement sur de petits effectifs et le plus souvent en modalité planaire. Le but de cette étude est d’évaluer ses reproductibilités intra- et inter-observateurs en modalités planaire et tomographique dans la détermination préopératoire du futur foie restant (FFR), l’influence de l’excrétion biliaire et du choix du seuil d’isocontour.

Patients et méthode

Chaque examen était traité deux fois par trois observateurs pour définir des ROI foie total (FT) et FFR en modalités planaire et tomoscintigraphique, après recalage avec le scanner injecté, en utilisant un isocontour par seuillage. La part du FFR était ensuite calculée, exprimée en pourcentage. Les reproductibilités intra- et inter-observateurs étaient calculées. Lorsque l’excrétion biliaire était visible, elle était corrigée et les résultats comparés avec les résultats initiaux. La variation du seuil d’isocontour (−4 %, −2 %, +2 % et +4 % par rapport au seuil initial) a aussi été étudiée.

Résultats

Trente patients ont été inclus. La reproductibilité intra-observateur était bonne (R2=0,847, 0,830 et 0,966) en acquisition planaire et meilleure en tomoscintigraphie (R2=0,957, 0,962 et 0,984). Il n’était mis en évidence aucune différence significative après correction des voies biliaires, quel que soit l’observateur (p=0,29, p=0,99 et p=0,61). Le changement de seuil entraînait des différences significatives mais minimes (1 % en moyenne) pour les seuils évalués.

Conclusion

La détermination du FFR par scintigraphie hépato-biliaire (planaire et tomoscintigraphique) est un examen reproductible, permettant d’envisager son évaluation sur des études multicentriques.

Le texte complet de cet article est disponible en PDF.

Abstract

Purpose

Performances of 99mTc-mebrofenin hepatobiliary scintigraphy (HBS) are currently evaluated before major hepatic resection for early detection of liver failure in high risk patients. Published studies included small numbers of patients with planar scintigraphy, and more recently SPECT data. Our aim was to evaluate intra- and interobserver reproducibility of HBS in planar and SPECT modalities for functional assessment of the future remnant liver (FRL) before major hepatic resection. We also evaluated the impact of potential variability factors such as biliary excretion and isocontour threshold choice.

Methods

All the scans were processed twice by three observers. Two regions of interest were drawn around total liver (TL) and FRL, on the planar dataset and with a threshold based algorithm on SPECT after registration with contrast-enhanced CT. Intra- and interobserver reproducibility of the FRL/TL ratio were calculated. When biliary excretion was already visible on SPECT acquisition, it was corrected and the ratio compared before and after correction. The results were also compared between different thresholds of the maximum liver value (−4%, −2%, +2%, +4% of the optimal threshold chosen by the observer).

Results

Thirty patients were included. The intraobserver reproducibility was good (R2=0.847, 0.830 and 0.966) when calculated on planar modality and better with SPECT (R2=0.957, 0.962 and 0.984). There was no significant difference after biliary excretion correction (P=0.29, P=0.99 and P=0.61 for the three observers). The threshold modification caused significant but minor differences (1% on average) for the evaluated thresholds.

Conclusions

The FRL functional evaluation with HBS including a SPECT acquisition provides a reproducible assessment, hardly affected by potential variability factors, allowing its use in multicentric studies.

Le texte complet de cet article est disponible en PDF.

Mots clés : Scintigraphie hépato-biliaire, 99mTc-mébrofénine, Reproductibilité, Hépatectomie, Quantification

Keywords : Hepatobiliary scintigraphy, 99mTc-mebrofenin, Reproducibility, Hepatectomy, Quantification


Plan


© 2015  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 39 - N° 2

P. 182-191 - avril 2015 Retour au numéro
Article précédent Article précédent
  • Intérêt de la quantification absolue du débit sanguin myocardique dans la sarcoïdose cardiaque par la TEP au 82-Rubidium
  • Q. Gillebert, R. Chequer, R. Ben Azzouna, S. Leygnac, N. Mikail, F. Hyafil, C. Chapelon-Albric, B. Crestani, D. Le Guludec, F. Rouzet
| Article suivant Article suivant
  • Recherche de contamination atmosphérique dans un service de médecine nucléaire
  • P. Damien, V. Guilloux, F. Blanc-Béguin, P. Guillo, P.Y. Salaün

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’achat d’article à l’unité est indisponible à l’heure actuelle.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.