Calibration of a Novel, Laparoscopic, 12-mm, Ultrasound, Image-guided, High-intensity Focused Ultrasound Probe for Ablation of Renal Neoplasms - 28/03/15

Doi : 10.1016/j.urology.2014.10.063 
Jason C. Sea a, Clinton D. Bahler a, Joshua D. Ring a, Sable Amstutz a, Narendra T. Sanghvi b, Liang Cheng c, Chandru P. Sundaram a,
a Department of Urology, Indiana University, Indianapolis, IN 
b SonaCare Medical, Indianapolis, IN 
c Department of Pathology, Indiana University, Indianapolis, IN 

Address correspondence to: Chandru P. Sundaram, M.D., Department of Urology, Indiana University, 535 N Barnhill Dr, Suite 420, Indianapolis, IN 46202.

Abstract

Objective

To calibrate a 12-mm high-intensity focused ultrasound (HIFU) probe to produce histologically verifiable complete tissue necrosis through thermal energy delivery while maintaining safety in ablation of porcine renal tissue.

Methods

The study was approved by the animal research committee at our institution. A HIFU probe on a fixed arm was introduced, and ablation was subsequently carried out robotically (under computer control) at 2 planned locations within each kidney at varying energy levels. After 4 days of survival, kidneys were retrieved and pathologic analysis of the lesions was carried out for necrotic volume. Planned volume was compared with pathologic necrotic volume.

Results

A series of optimizations were carried out as the experiment progressed. Near optimal ablation zone dimensions with complete cell death were achieved when the energy density was between 400 and 460 calories/cm3. Pathologic analysis demonstrated the average ratio of necrotic zone volume to planned zone volume was 1.0 (standard deviation, 0.18) in the 7 lesions with energy density between 400 and 460 calories/cm3. All animals survived the surgeries, and no device-related complications were identified during the study. No acute renal failure was identified in analysis of the blood work.

Conclusion

A novel 12-mm laparoscopic HIFU probe was successfully calibrated for the ablation of porcine renal tissue. The safety profile of the Sonatherm probe was found to be excellent with no complications identified.

Le texte complet de cet article est disponible en PDF.

Plan


 Financial Disclosure: Narendra T. Sanghvi is an employee of SonaCare Medical. The remaining authors declare that they have no relevant financial interests.
 Funding Support: This study was supported by SonaCare Medical.


© 2015  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 85 - N° 4

P. 953-958 - avril 2015 Retour au numéro
Article précédent Article précédent
  • Microsurgical Varicocelectomy With Transfixing of the Difficult-to-isolate Periarterial Vein Using Microsutures
  • Yan Zhang, Xiaojian Yang, Xiao Wu, Hao Zhang, Shitao Chen, Xin Gao
| Article suivant Article suivant
  • Transition of Urologic Patients From Pediatric to Adult Care: A Preliminary Assessment of Readiness in Spina Bifida Patients
  • Heidi A. Stephany, Christina B. Ching, Melissa R. Kaufman, Amanda Squiers, Lisa Trusler, Douglass B. Clayton, John C. Thomas, John C. Pope, Mark C. Adams, John W. Brock, Stacy T. Tanaka

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.