Usefulness and Safety of Ultrasound-Assisted Catheter-Directed Thrombolysis for Submassive Pulmonary Emboli - 28/02/15
Abstract |
The optimal treatment for intermediate-risk pulmonary embolism (PE) remains unclear. Our goal was to describe the safety and efficacy of the EkoSonic ultrasound-assisted catheter-directed thrombolysis system (EKOS Corporation, Bothell, Washington) in a real-world registry of patients with intermediate-risk PE. Fifty-three consecutive patients with intermediate-risk PE treated with ultrasound-assisted catheter-directed thrombolysis at Brigham and Women's Hospital from 2010 to 2014 were analyzed. The primary outcome was a change in directly measured pulmonary artery pressures as assessed using logistic regression with generalized estimating equations to account for serial measurements. Patients received an average of 24.6 ± 9 mg of alteplase using the EKOS catheter with an average treatment time of 15.9 ± 3 hours. After treatment, there was a 7.2- and a 11.4-mm Hg reduction in mean and systolic pulmonary artery pressure (95% confidence interval 4.7 to 9.7 mm Hg, p <0.001, and 95% confidence interval 7.8 to 15.0 mm Hg, p <0.001), respectively. In this cohort, 9.4% had any bleeding complication noted during their hospital stay. One patient's alteplase was prematurely discontinued for access site bleeding although no other interventions were required related to bleeding complications.
Le texte complet de cet article est disponible en PDF.Plan
See page 824 for disclosure information. |
Vol 115 - N° 6
P. 821-824 - mars 2015 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.
Déjà abonné à cette revue ?