S'abonner

Pseudo-malfunction of the Coloplast Titan Inflatable Penile Prosthesis One-Touch Release Pump - 26/09/14

Doi : 10.1016/j.urology.2014.06.012 
Bruce B. Garber a, Jacob L. Khurgin b, , Doron S. Stember c, Paul E. Perito d
a Department of Urology, Hahnemann University Hospital, Drexel University College of Medicine, Philadelphia, PA 
b Department of Urology, Einstein Medical Center, Philadelphia, PA 
c Department of Urology, Mount Sinai Beth Israel, New York, NY 
d Department of Urology, Coral Gables Hospital, Coral Gables, FL 

Reprint requests: Jacob L. Khurgin, D.O., Einstein Medical Center, 5501 Old York Rd, Klein 500, Philadelphia, PA 19141.

Abstract

Objective

To define and describe a type of pseudo-malfunction of the Coloplast Titan Inflatable Penile Prosthesis (IPP) One-Touch Release (OTR) pump (Coloplast Corp, Minneapolis, MN).

Materials and Methods

We retrospectively reviewed a consecutive series of 550 patients with refractory organic erectile dysfunction who were implanted with a Coloplast Titan IPP with the OTR pump during a period of approximately 4 years.

Results

All patients were implanted using standard techniques through an infrapubic or penoscrotal approach. Twenty-nine patients (5.3%) complained that their IPP would not inflate and that the pump bulb felt “hard.” Examination revealed that their IPP was working normally; however, the inflate/deflate valve disc had become stuck in the deflate position. Very firm pressure had to be applied to the pump bulb to move the valve disc into the inflate position. Once this was accomplished, the device inflated and deflated normally. Another 14 patients (2.5%) reported this phenomenon to us but were able to apply enough pressure on the pump bulb to rectify it.

Conclusion

The inflate/deflate valve disc in the Coloplast Titan OTR pump can occasionally become stuck in the deflate position (7.8% of patients in our experience). Patients may be unable to inflate the device and return for evaluation. In all cases we have encountered, firm pressure on the pump bulb caused the valve to shift into the inflate position, and the device worked properly thereafter. Patients and implanting urologists should be aware of this issue and of the way in which it can be rectified.

Le texte complet de cet article est disponible en PDF.

Plan


 Financial Disclosure: Bruce B. Garber and Paul E. Perito are paid consultants to Coloplast and American Medical Systems. The remaining authors declare that they have no relevant financial interests.


© 2014  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 84 - N° 4

P. 857-859 - octobre 2014 Retour au numéro
Article précédent Article précédent
  • Withdrawn: “Sulforaphane Restores Oxidative Stress Induced by Di-n-butylphthalate in Testicular Leydig Cells With Low Basal Reactive Oxygen Species Levels” [URL 84 (2014) 850 - 856]
  • Baixin Shen, Wei Wang, Long Ma, Shangqian Wang, Liucheng Ding, Zhengsen Chen, Yunpeng Sao, Hua Shen, Zhongqing Wei, Wei Zhang
| Article suivant Article suivant
  • Who Really Benefits From Nephron-sparing Surgery?
  • Solomon L. Woldu, Aaron C. Weinberg, Ruslan Korets, Rashed Ghandour, Matthew R. Danzig, Arindam RoyChoudhury, Sean D. Kalloo, Mitchell C. Benson, G. Joel DeCastro, James M. McKiernan

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2025 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.