Gemstone Spectral Imaging Dual-energy Computed Tomography: A Novel Technique to Determine Urinary Stone Composition - 27/03/13

Doi : 10.1016/j.urology.2013.01.010 
Xiaohu Li a, Ren Zhao b, Bin Liu a, , Yongqiang Yu a,
a Department of Radiology, First Affiliated Hospital of Anhui Medical University, Hefei, China 
b Department of Cardiology, First Affiliated Hospital of Anhui Medical University, Hefei, China 

Reprint requests: Bin Liu, M.M., and Yongqiang Yu, M.D., Department of Radiology, First Affiliated Hospital of Anhui Medical University, No. 218 Jixi Er Road, Hefei 230022, China.

Abstract

Objective

To assess the capability of determining the chemical composition of urinary stones using dual-energy spectral computed tomography (CT).

Materials and Methods

A total of 116 urinary stones with a known chemical composition determined by infrared spectroscopy were scanned using both gemstone spectral imaging and conventional polychromatic imaging (at 120 kVp) in vitro. The CT numbers of each stone were recorded from the 11 sets of monochromatic images (40-140 keV, 10-keV interval) and 120-kVp polychromatic images. The differences between the CT numbers obtained in the 12 sets were compared among the stone groups.

Results

The 116 urinary stones were classified into 7 groups: uric acid (n = 16), cystine (n = 10), brushite (n = 17), weddellite (n = 14), whewellite (n = 24), carbapatite (n = 12), and struvite (n = 23). The paired CT numbers (at 120 kVp and 50 keV, respectively) of each stone group were as follows: uric acid, 469.11 ± 69.83 and 474.06 ± 40.55 Hounsfield units (HU); cystine, 564.81 ± 97.71 and 839.58 ± 92.14 HU; brushite, 1830.94 ± 59.19 and 2787.41 ± 306.07 HU; weddellite, 1438.09 ± 191.12 and 2100.79 ± 202.32 HU; whewellite, 1493.48 ± 259.90 and 2321.39 ± 367.23 HU; carbapatite, 1784.58 ± 106.42 and 2513.86 ± 189.09 HU; and struvite, 833.037 ± 9.91 and 1123.24 ± 267.70 HU. The differences in the CT numbers at 120 kVp and 50 keV among the groups were statistically significant by binary comparison (P <.05), except for those at 120 kVp between uric acid and cystine (P = .121), whewellite and weddellite (P = .280), and brushite and carbapatite (P = .419).

Conclusion

Gemstone spectral imaging dual-energy CT provides a novel method to better characterize pure urinary stones using the CT numbers at 50 keV.

Le texte complet de cet article est disponible en PDF.

Plan


 Xiaohu Li and Ren Zhao contributed equally to the present report.
 Financial Disclosure: The authors declare that they have no relevant financial interests.


© 2013  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 81 - N° 4

P. 727-730 - avril 2013 Retour au numéro
Article précédent Article précédent
  • Clinically Insignificant Residual Fragments: An Acceptable Term in the Computed Tomography Era?
  • Yasser Osman, Ahmed M. Harraz, Ahmed R. El-Nahas, Bassam Awad, Naser El-Tabey, Haytham Shebel, Ahmed M. Shoma, Ibrahim Eraky, Mahmoud El-Kenawy
| Article suivant Article suivant
  • Effects of Aminobisphosphonates and Thiazides in Patients With Osteopenia/Osteoporosis, Hypercalciuria, and Recurring Renal Calcium Lithiasis
  • Miguel Angel Arrabal-Polo, Salvador Arias-Santiago, Tomas de Haro-Muñoz, Adriana Lopez-Ruiz, Jacinto Orgaz-Molina, Samuel Gonzalez-Torres, Armando Zuluaga-Gomez, Miguel Arrabal-Martin

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.