Évolution de la pression digitale pendant une séance de dialyse chez 49 patients - 29/11/12

Doi : 10.1016/j.jmv.2012.10.002 
J. Journet a, , C. Barbe b, P. Rieu c, H. Maheut c, K. Gaha c, I. Kazes c, A. Long a
a Service de médecine vasculaire, hôpital Robert-Debré, centre hospitalier universitaire de Reims, avenue du Général-Koenig, 51092 Reims cedex, France 
b Unité d’aide méthodologique, hôpital Robert-Debré, centre hospitalier universitaire de Reims, avenue du Général-Koenig, 51092 Reims cedex, France 
c Service de néphrologie, hôpital Maison-Blanche, centre hospitalier universitaire de Reims, 45, rue Cognacq-Jay, 51092 Reims cedex, France 

Auteur correspondant.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Résumé

Objectifs

Décrire et comparer l’évolution de la pression digitale aux deux mains au cours d’une séance de dialyse chez des patients sans symptomatologie d’ischémie digitale et identifier les paramètres influençant la pression digitale.

Patients

Les patients hémodialysés sur un abord vasculaire au membre supérieur étaient inclus de façon prospective. La pression systolique digitale du troisième doigt des deux mains mesurée par photopléthysmographie (PPG) et la pression systolique brachiale étaient recueillies avant la dialyse (H0) puis toutes les heures (de H1 à H4).

Résultats

Sur 53 patients, 49 ont été inclus (exclusions : une chirurgie pour ischémie, un tremblement distal, deux absences de consentement). Aucun ne présentait d’ischémie digitale. La pression digitale était significativement inférieure du côté de l’abord vasculaire par rapport au côté controlatéral avant et pendant la dialyse. La pression digitale diminuait significativement des deux côtés pendant la dialyse. La pression brachiale diminuait significativement pendant la dialyse. Seule la diminution de la pression brachiale était corrélée à la diminution des pressions digitales. La pression digitale était inférieure à 30mmHg chez six patients. Aucun n’a développé d’ischémie digitale après six mois de suivi.

Conclusion

À notre connaissance, cette étude est la première à montrer une diminution significative des pressions digitales en cours de dialyse aux deux mains chez des patients sans symptomatologie d’ischémie digitale. D’autres travaux sont nécessaires afin d’identifier les paramètres pouvant influencer la pression digitale ainsi que l’utilité de cette mesure en pratique courante.

Le texte complet de cet article est disponible en PDF.

Summary

Objectives

To describe and compare the evolution of digital pressure on both hands during a dialysis session in patients without digital ischemia, and to identify the parameters influencing the digital pressure.

Materials

Patients with an upper limb vascular access were prospectively included. Digital systolic pressure on the third finger of both hands measured by photoplethysmography and brachial systolic pressure were recorded before dialysis (H0) and every hour (from H1 to H4).

Results

Among 53 patients, 49 were included (exclusions: one surgery for ischemia, one hand tremor, two no consent). None of them had digital ischemia. Digital pressure homolateral to the vascular access was significantly lower compared with controlateral side before and during dialysis. Digital pressure significantly decreased on both sides during dialysis. Brachial pressure decreased significantly compared to H0. Only the brachial pressure decrease was correlated with the decrease of digital pressure. The digital pressure was less than 30mmHg in six patients. No evidence of digital ischemia was reported after a 6-month follow-up.

Conclusion

To our knowledge, this is the first study showing a significant decrease of digital pressure in both hands during hemodialysis in patients without digital ischemia. Further studies are necessary to investigate which parameters can affect digital pressure and to look for clinical consequence of this measurement.

Le texte complet de cet article est disponible en PDF.

Mots clés : Hémodialyse, Fistule artério-veineuse, Pression artérielle digitale

Keywords : Hemodialysis, Arteriovenous fistula, Digital/blood supply


Plan


 Ce travail a été présenté sous la forme d’une communication orale au Collège français de pathologie vasculaire, à Paris le 15 mars 2012.


© 2012  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 37 - N° 6

P. 283-290 - décembre 2012 Retour au numéro
Article précédent Article précédent
  • Hypertension artérielle : des objectifs pour 2015, une initiative originale
  • J. Blacher
| Article suivant Article suivant
  • Les paragangliomes cervicaux. Une expérience chirurgicale à propos de 12 cas
  • Y. Tijani, H. Chtata, B.T. Kouvidjin, Y. Abissegue, M. Taberkant

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.