Is ovarian volume estimation reliable when compared with true volume? - 23/12/11
Résumé |
Objective |
We aimed to evaluate the agreement of 2-dimensional (2D) and 3-dimensional (3D) ultrasonography (USG) with true ovarian volume (OV), as calculated precisely after oophorectomy.
Study Design |
A total of 46 ovaries from 30 patients were prospectively enrolled. Preoperatively, all ovaries were assessed by 2D and 3D USG for volume estimation and results were compared with true OV that was calculated with Archimedes' principles following oophorectomy.
Results |
The correlation coefficients of 2D and 3D USG with true OV were similar (0.65 vs 0.67, respectively). The mean bias (upper and lower limits of agreement) between 2D and true OV was 1.41 (–3.84 to 6.66) mL. The respective figure for 3D and true OV were 0.33 (–4.71 to 5.37) mL. While estimation by 2D USG brought 18% larger, 3D USG revealed 11% smaller values than the true OV.
Conclusion |
Three-dimensional OV estimation might present improvement in means of lower mean bias than 2D USG.
Le texte complet de cet article est disponible en PDF.Key words : agreement, correlation, formula, ovarian volume, ultrasonography
Plan
The authors report no conflict of interest |
|
Cite this article as: Bozdag G, Salman MC, Mumusoglu S, et al. Is ovarian volume estimation reliable when compared with true volume? Am J Obstet Gynecol 2012;206:44.e1-4. |