A novel hands-free carotid ultrasound detects low-flow cardiac output in a swine model of pulseless electrical activity arrest - 10/11/11
Abstract |
Objective |
To determine if a hands-free, noninvasive Doppler ultrasound device can reliably detect low-flow cardiac output by measuring carotid artery blood flow velocities. We compared the ability of observers to detect carotid artery flow velocity differences between pseudo-pulseless electrical activity (PEA) and true-PEA cardiac arrest.
Methods |
Five swine were instrumented with aortic (Ao) and right atrial pressure-transducing catheters. The Doppler ultrasound device was adhered to the neck over the carotid artery. Continuous electrocardiogram, pressure readings, and Doppler signal were recorded. Each swine underwent multiple episodes of fibrillation and resuscitation. Episodes of true-PEA and pseudo-PEA were retrospectively identified from all resuscitation attempts by examination of electrocardiogram and Ao waveforms. The sensitivity and specificity of the device to detect pseudo-PEA was obtained using observers blinded to Ao waveform recordings.
Results |
There was good interobserver reliability related to identification of pseudo- and true-PEA (κ = 0.873). The observers blinded to Ao waveform recordings agreed on 8 of the 9 episodes of pseudo-PEA, whereas 4 false positives of 26 true-PEA events were reported (sensitivity, 0.89; specificity, 0.85). The Doppler device was able to detect carotid flow velocity over a wide range of Ao blood pressures.
Conclusions |
This hands-free, noninvasive Doppler ultrasound device can reliably differentiate pseudo-PEA from true-PEA during resuscitation from cardiac arrest, detecting pressure gradient changes of less than 5 mm Hg through to normotension. This device distinguishes conditions of no cardiac output from low cardiac output and may have applications for use during resuscitation from various etiologies of arrest and shock.
Le texte complet de cet article est disponible en PDF.Plan
☆ | Financial support: This work was supported by Philips Medical Systems, Andover, Mass. |
☆☆ | Presented as an abstract at the American Heart Association Resuscitation Sciences Symposium, Chicago, Ill, November, 2006 [Circulation (supplement) 2006. 114(18): II-1207]. |
★ | Conflict of interest statement: This study was funded by Philips Medical Systems. Dr Larabee and Dr Little received financial support for the performance of this study and have no financial interest in the device. All other investigators are employed by Philips Medical Systems. |
Vol 29 - N° 9
P. 1141-1146 - novembre 2011 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.
Déjà abonné à cette revue ?