Measuring pressure-derived fractional flow reserve through four french diagnostic catheters - 28/08/11

Doi : 10.1016/S0002-9149(03)00151-6 
Pierre Legalery, MD a, Marie-France Seronde, MD a, Nicolas Meneveau, MD, PhD a, François Schiele, MD, PhD a, , Jean-Pierre Bassand, MD a
a University Hospital Jean-Minjoz, Besançon, France 

*Address for reprints: François Schiele, MD, PhD, Department of Cardiology, University Hospital Jean-Minjoz, Boulevard Fleming, 25030 Besançon Cedex, France.

Abstract

Measurement of fractional flow reserve (FFR) with a pressure wire is used to discriminate between patients with and without functionally significant lesions. FFR can be assessed through a conventional 4Fr diagnostic catheter, which is a more convenient method of assessment. The aim of this study was to assess the feasibility, safety, repeatability, and reliability of routine FFR measurements through 4Fr diagnostic catheters. From a single-center prospective registry, results of FFR assessment through a 4Fr catheter were used to determine: (1) feasibility and safety, by the procedural success rate and immediate and 30-day clinical outcome; (2) repeatability, by the intraclass correlation coefficient and comparison of the difference (means ± 2 SDs); and (3) reliability, by comparison of results obtained using 4Fr versus 7Fr guiding catheters. During the study period, FFR was measured in 190 patients, in 122 using a diagnostic 4Fr catheter (study population) and in 68 using a 7Fr guiding catheter. Measurement of FFR wa successful in 115 of 122 patients (94%). No complications related to the use of the 4Fr catheter occurred. Repeatability was determined from 108 repeated measurements: the intraclass correlation coefficient was 0.942 and the mean difference between repeated FFR measurements was −0.001 ± 0.038. Reliability was determined from 15 unselected patients; there was no systematic error and only 1 value was out of the range of 2 SDs of the mean difference. Using a threshold value of 0.75, the Kappa coefficient for the qualitative agreement was 0.84. Thus, pressure-derived FFR assessment can safely be performed through 4Fr diagnostic catheters, with similar repeatability and reliability as 7Fr guiding catheters, resulting in a simplification of the measurement procedure.

Le texte complet de cet article est disponible en PDF.

Plan


© 2003  Excerpta Medica Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 91 - N° 9

P. 1075-1078 - mai 2003 Retour au numéro
Article précédent Article précédent
  • Impact of minor electrocardiographic ST-segment and/or T-wave abnormalities on cardiovascular mortality during long-term follow-up
  • Philip Greenland, Xiaoyuan Xie, Kiang Liu, Laura Colangelo, Youlian Liao, Martha L Daviglus, Abby N Agulnek, Jeremiah Stamler
| Article suivant Article suivant
  • Doppler transmitral flow indexes and risk of atrial fibrillation (The Framingham Heart Study)
  • Ramachandran S Vasan, Martin G Larson, Daniel Levy, Maurizio Galderisi, Philip A Wolf, Emelia J Benjamin

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.