Usefulness of a hand-carried cardiac ultrasound device to detect clinically significant valvular regurgitation in hospitalized patients - 26/08/11
Abstract |
The accuracy of a hand-carried cardiac ultrasound (HCCU) device for the diagnosis of valvular regurgitations using color-flow Doppler was assessed. One hundred twenty hospitalized patients with at least mild valvular regurgitation by standard echocardiography were evaluated by 2 cardiologists using HCCU. The HCCU operator identified 99% of clinically significant valvular regurgitations and assessed the severity correctly in 83% of cases. For mild regurgitation, the HCCU sensitivity and specificity were 82% and 93%, respectively, resulting in correct assessments of severity in 71% of cases. HCCU used by trained cardiologists has high sensitivity for the detection and accurate assessment of the severity of clinically relevant valvular regurgitation.
Le texte complet de cet article est disponible en PDF.Vol 93 - N° 8
P. 1069-1072 - avril 2004 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.