Effect of Repetitive Loading on the Mechanical Properties of Synthetic Hernia Repair Materials - 24/08/11

Doi : 10.1016/j.jamcollsurg.2011.05.018 
Braden J. Eliason, Margaret M. Frisella, RN, Brent D. Matthews, MD, FACS, Corey R. Deeken, PhD
Department of Surgery, Section of Minimally Invasive Surgery, Washington University School of Medicine, St Louis, MO 

Correspondence address: Corey R Deeken, PhD, Department of Surgery, Washington University School of Medicine, 660 South Euclid Ave, Campus Box 8109, St Louis, MO 63110

Résumé

Background

Hernia repair materials undergo repeated loading while in the body, and the impact on mechanical properties is unknown. It was hypothesized that exposure to repetitive loading would lead to decreased tensile strength and increased strain, and that these differences would become more pronounced with greater loading and unloading sequences.

Study Design

Polypropylene, expanded polytetrafluoroethylene, composite barrier, and partially absorbable meshes were evaluated. Twenty specimens (7.5 × 7.5 cm) were prepared from each material. Five specimens were subjected to ball burst testing to determine baseline biomechanical properties. Cycles of 10, 100, and 1,000 loading sequences were also performed (n = 5 each).

Results

BardMesh (CR Bard/Davol), Dualmesh (WL Gore), and Prolene (Ethicon) exhibited significantly reduced tensile strength; BardMesh, Proceed (Ethicon), Prolene, ProLite (Atrium Medical), ProLite Ultra (Atrium Medical), and Ultrapro (Ethicon) exhibited significantly increased strain after exposure to 1,000 cycles compared with their baseline properties. BardMesh and Prolene demonstrated both reduced tensile strength and increased strain values after 1,000 cycles, suggesting that repetitive loading has the greatest effects on these materials. In addition, BardMesh and Prolene exhibited progressively worsening effects as the number of cycles was increased.

Conclusions

Deterioration of the tensile strength of the mesh or an increase in the ability of the mesh material to stretch (ie, increased strain values) could potentially lead to hernia recurrence or a poor functional result. However, the results of this study should not be interpreted to mean that hernia repair materials will fail in the body after only 10, 100, or 1,000 cycles. The conditions used in this study were more extreme than most physiologic scenarios and were intended as a pilot investigation into how the mechanical properties of hernia repair materials are affected by in vitro cyclic testing.

Le texte complet de cet article est disponible en PDF.

Plan


 Disclosure Information: Dr Deeken is a consultant for Atrium Medical Corporation and Davol Incorporated. Dr Matthews is a consultant for Atrium Medical Corporation and Ethicon Incorporated. He also receives honoraria and research and equipment support from Atrium Medical Corporation, Ethicon EndoSurgery, Karl Storz Endoscopy, Stryker Endoscopy, and WL Gore & Associates, Incorporated. All other authors have nothing to disclose.
 Funding for this project was provided by the Washington University Institute for Minimally Invasive Surgery (WUIMIS, St Louis, MO).


© 2011  American College of Surgeons. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 213 - N° 3

P. 430-435 - septembre 2011 Retour au numéro
Article précédent Article précédent
  • A Novel Technique for Natural Orifice Endoscopic Full-Thickness Colon Wall Resection: An Experimental Pilot Study
  • Erwin Rieder, Danny V. Martinec, Christy M. Dunst, Lee L. Swanström
| Article suivant Article suivant
  • Current Treatment for the Local Control of Retroperitoneal Sarcomas
  • Edmund Bartlett, Sam S. Yoon

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.