Evaluation of Lens Epithelial Cell Migration In Vivo at the Haptic-Optic Junction of a One-Piece Hydrophobic Acrylic Intraocular Lens - 18/08/11
Résumé |
Purpose |
To evaluate the pattern of lens epithelial cell (LEC) ingrowth behind the intraocular lens (IOL) optic in patients implanted with the AcrySof SA60AT one-piece IOL.
Design |
Retrospective case series.
Methods |
Patients implanted with the AcrySof SA60AT one-piece IOL with complete overlap of the capsulorrhexis and IOL optic and no ocular pathology were selected from the practices of six surgeons. High-resolution digital images of the distribution of LECs were captured. Circumferential location and extent of LEC growth behind the optic were measured.
Results |
Mean ± SD follow-up was 13.2 ± 1.7 months for 40 patients (mean age 72 years). LEC migration beyond the edge of the IOL optic was displayed by 57.5% (23/40) of patients. LECs migrated into the visual axis in 22.5% (9/40) of patients, leading to a visual acuity of <6/9 (20/30) in 10% (4/40) of patients. Cell migration was located preferentially at the optic-haptic junction. Serial photographs were used to monitor LEC migration over time. In some cases, LECs extended from the opposing optic-haptic junctions to converge in the center of the IOL, creating a path across the optic.
Conclusions |
The optic-haptic junction of the AcrySof one-piece IOL is a point of weakness in the barrier effect of the square-edge IOL design that provides migrating LECs access to the posterior capsule.
Le texte complet de cet article est disponible en PDF.Plan
Supported in part by an unrestricted grant from Advanced Medical Optics Inc. |
Vol 142 - N° 4
P. 557-562 - octobre 2006 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.
Déjà abonné à cette revue ?