In-line radiofrequency ablation to minimize blood loss in hepatic parenchymal transection - 18/08/11
Abstract |
Background |
Intraoperative blood loss has been shown to be an important factor correlating with morbidity and mortality in liver surgery. A 5-cm long instrument with variably deployable metal electrodes using in-line radiofrequency ablation (ILRFA) energy was used for hepatic transection in an attempt to reduce bleeding.
Methods |
Eight patients underwent liver resection. At each resection, half the resection was performed with ILRFA and the other half was performed with an ultrasonic aspirator alone. Blood loss was measured for each mode of resection.
Results |
The mean blood loss using ILRFA was 6.5 (±3.7) mL/cm2 compared with 20.4 (±8.7) mL/cm2 by using the ultrasonic aspirator (P = .004).
Conclusions |
In-line radiofrequency ablation reduced bleeding during hepatic parenchymal transection when compared with the ultrasonic aspirator.
Le texte complet de cet article est disponible en PDF.Keywords : Radiofrequency ablation, Liver resection, Hemorrhage, Disclosure
Plan
The InLine probe is now owned by Resect Medical. D.L.M. and S.D. declare a financial interest in this company. |