S'abonner

Mid-term Results of the Hancock II Valve and Carpentier-Edward Perimount Valve in the Pulmonary Portion in Congenital Heart Disease - 11/08/11

Doi : 10.1016/j.hlc.2009.10.009 
Jae Gun Kwak, MD a, Jeong Ryul Lee, MD b, , Woong-Han Kim, MD b, Yong Jin Kim, MD b
a Department of Thoracic and Cardiovascular Surgery, Sejong General Hospital, Bucheon, Republic of Korea 
b Department of Thoracic and Cardiovascular Surgery, Seoul National University Children’s Hospital, 28 Yeongun Dong, Jongno Gu, Seoul, Republic of Korea 

Corresponding author. Tel.: +82 11 9989 1081.

Résumé

Background

As the number of cases with artificial pulmonary valve implantation increases for congenital heart disease, the number of young adults with artificial pulmonary valves has also increased.

Methods

From 2000 to 2007, 146 artificial valves, such as the Carpentier-Edward Perimount, Hancock II, Biocor, homograft and hand-made valves were implanted for pulmonary valve in 132 patients with various forms of congenital heart disease. Among them, the outcomes of the Carpentier-Edward Perimount (n=63) and the Hancock II (n=40) valves were reviewed retrospectively. The mean age at initial implantation was 12.8±6.6 years. The overall duration of follow up was 36.0±24.2 months.

Results

There was an early death due to right ventricular failure with intractable ventricular arrhythmia and 3 late deaths due to progressive right ventricular failure, dilated cardiomyopathy and infective endocarditis. The overall survival and re-operation free rate was 96.3% and 89.8% respectively. Eight out of 63 Carpentier-Edward patients (12.6%) underwent re-replacement at 49.2±25.2 months. The re-operation free rates were 97.7%, 87.7% and 50% at 1, 3 and 5 years respectively. There was no re-operation required for the 40 Hancock II patients over 18.0±10.8 months. There was no statistical significance in the re-operation free rates between these 2 valves (p-value=0.51).

Conclusions

The overall survival rate associated with pulmonary valve bioprosthetic valve implantation was acceptable. However, the re-operation freedom rate was not satisfactory at mid-term for the Carpentier-Edward.

Le texte complet de cet article est disponible en PDF.

Keywords : Congenital heart disease, Heart valve, prosthesis, Heart valve, allograft, Pulmonary valve replacement


Plan


© 2009  Australasian Society of Cardiac and Thoracic Surgeons and the Cardiac Society of Australia and New Zealand. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 19 - N° 4

P. 243-246 - avril 2010 Retour au numéro
Article précédent Article précédent
  • Serum Thromboxane B2 Compared to Five Other Platelet Function Tests for the Evaluation of Aspirin Effect in Stable Cardiovascular Disease
  • Giselle Kidson-Gerber, James Weaver, Rosalie Gemmell, Ananth M. Prasan, Beng Hock Chong
| Article suivant Article suivant
  • The Use of Coronary Revascularisation Procedures in Urban Australian Aboriginals and a Matched General Population : Coronary Procedures in Aboriginals
  • Pamela J. Bradshaw, Helman S. Alfonso, Judith Finn, Julie Owen, Peter L. Thompson

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.