Thermostable adenylate kinase technology: a new process indicator and its use as a validation tool for the reprocessing of surgical instruments - 07/08/11

Doi : 10.1016/j.jhin.2009.07.019 
J.R. Hesp a, , , T.M. Poolman a, , C. Budge a, L. Batten a, F. Alexander a, G. McLuckie a, S. O'Brien a, P. Wells b, N.D.H. Raven a, J.M. Sutton a
a Technology Development Group, Centre for Emergency Response and Preparedness, Health Protection Agency, Porton Down, Salisbury, Wiltshire, UK 
b Sterilisation and Disinfection Unit, Salisbury NHS Foundation Trust, Salisbury District Hospital, Salisbury, Wiltshire, UK 

Corresponding author. Address: Technology Development Group, Health Protection Agency, Centre for Emergency Preparedness and Response, Porton Down, Salisbury, Wiltshire SP4 0JG, UK. Tel.: +44 (0) 1980 612952; fax: +44 (0) 1980 612622.

Summary

Adenylate kinase (tAK), a thermostable enzyme, was assessed as a possible means of providing a quantitative measure of cleaning efficacy suitable for validating the performance of an automated washer disinfector (AWD) during routine use. Two indicator formulations were developed using either a commercially available washer disinfector soil or a protein-based soil. Each indicator consisted of 100μg (in test soil) of tAK dried on to a steel or plastic surface. These indicators were placed in each basket of a washer disinfector and processed alongside soiled surgical instruments during a standard day's operation. After processing, remaining tAK activity was detected using a rapid enzyme assay (2min detection time) in a handheld hygiene monitor. The amount of tAK remaining on each indictor after a full AWD cycle was found to range from 0.1 to 0.4ng, which represented a mean log10 removal of 5.8±0.3. There was no statistical difference in the residual tAK activity between individual runs or the position of the indicator in the machine. The tAK indicator was also used to analyse the protein removal within each component of the wash cycle. These results demonstrated that all phases of the wash process contributed to the removal of the protein load, with the main wash alone being responsible for 3.6–4.0log10 reductions in protein activity. We propose that a quantitative cleaning index using such rapid readout indicator devices would provide a valuable addition to the methodologies for validating cleaning processes.

Le texte complet de cet article est disponible en PDF.

Keywords : Adenylate kinase, Cleaning process validation, Decontamination


Plan


© 2009  The Hospital Infection Society. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 74 - N° 2

P. 137-143 - février 2010 Retour au numéro
Article précédent Article précédent
  • Effect of anaesthesia maintained with sevoflurane and propofol on surgical site infection after elective open gastrointestinal surgery
  • K. Shimizu, M. Hirose, S. Mikami, K. Takamura, T. Goi, A. Yamaguchi, K. Morioka, T. Ichikawa, K. Shigemi
| Article suivant Article suivant
  • Quantitative measurement of the efficacy of protein removal by cleaning formulations; comparative evaluation of prion-directed cleaning chemistries
  • M. Ungurs, J.R. Hesp, T. Poolman, G. McLuckie, S. O'Brien, H. Murdoch, P. Wells, N.D.H. Raven, J.T. Walker, J.M. Sutton

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.