Rôle de la cytotoxicité du milieu de transport d’implants : cas de lentilles intraoculaires conditionnées en injecteur - 13/02/25

Role of conditioning medium cytotoxicity. An example with an intraocular lens preloaded in injector

Doi : 10.1016/S0181-5512(06)73846-3 
A. Staropoli 1, C. Baudouin 2, 3, M. Dutot 1, 2, G. Sultan 1, 3, J.-M. Warnet 1, 2, P. Rat 1, 2,
1 Laboratoire de Toxicologie, Faculté de Pharmacie, Université Paris 5, France 
2 INSERM U598, Paris, France 
3 Service d’Ophtalmologie III, Centre Hospitalier National d’Ophtalmologie des Quinze-Vingts, Paris, France 

Correspondance : P. Rat, Laboratoire de Toxicologie (INSERM U598), Faculté de Pharmacie (Université Paris 5), 4, avenue de l’Observatoire, 75006 Paris.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Résumé

Introduction

À la suite de cas de matériovigilance rencontrés avec l’implant ACR6D SE® conditionné en injecteur Premier®, nous avons étudié la cytotoxicité de la lentille intraoculaire et de son conditionnement afin d’identifier l’élément responsable et de proposer des solutions mieux adaptées.

Matériels et méthodes

Des tests de biocompatibilité-cytotoxicité ont été réalisés selon les recommandations de la norme ISO 10993-5 sur la lignée de cellules épithéliales cristalliniennes humaines SRA 01/04. La réponse cellulaire a été étudiée grâce à l’utilisation de sondes fluorescentes : le Rouge neutre qui mesure l’intégrité membranaire, le Hoechst 33342 qui mesure l’ADN total, ainsi qu’une sonde qui étudie la fluidité membranaire, le YO-PRO-1. Les échantillons solides (lentille et injecteur) ont été testés après élution (norme ISO 10993-5). Les échantillons liquides (liquides de conditionnement, solutions salines) étaient testés purs directement sur les cellules.

Résultats

La lentille intraoculaire et le dispositif d’injection n’induisent pas de cytotoxicité dans les conditions de notre étude. En revanche, le liquide de conditionnement induit une perturbation de la fluidité membranaire (apoptose/inflammation). Cette modification pourrait expliquer la survenue de complications postopératoires. Les tests réalisés en insu sur les nouvelles versions d’implants ont permis d’identifier un nouveau liquide de conditionnement mieux adapté.

Conclusion

Nos résultats ont montré le rôle du liquide de conditionnement dans les problèmes de biocompatibilité-cytotoxicité du dispositif médical et dans la survenue de complications postopératoires. Le changement du liquide de conditionnement (milieu à ionisation contrôlée, riche en cations divalents) a permis la suppression des marqueurs biologiques de cette cytotoxicité.

Le texte complet de cet article est disponible en PDF.

Abstract

Introduction

Following material vigilance cases encountered with the hydrophilic acrylic intraocular lens, ACR6D SE preloaded in the Premier® shooter, we studied the cytotoxicity of the intraocular lens and its conditioning to identify the cytotoxic element. We proposed medical device modification to improve its biocompatibility.

Materials and Methods

Biocompatibility-cytotoxicity assays were carried out according to ISO 10993-5 recommendations. Tests were performed on the SRA 01/04 human lens epithelial cell line. Neutral red, Hoechst 33342, and YO-PRO-1 fluorescent probes were used to assess membrane integrity, total DNA, and membrane fluidity, respectively. Materials samples were prepared in culture medium according to the ISO 10993-5 elution procedure. Pure saline solutions and conditioning liquids were tested directly on cells.

Results

The intraocular lens and injector were not cytotoxic. Conditioning liquids induced membrane fluidity perturbation characteristic of apoptosis. Tests performed on new versions of the medical device identified a better adapted conditioning liquid.

Conclusion

The results suggest that the cytotoxicity of the conditioning liquid could explain the postoperative complication rate. When we changed the conditioning liquid with sterile irrigating solution (i.e., rich divalent cation marine solution), we eliminated cellular stress. Fluorescent probes are well adapted to assess medical device biocompatibility-cytotoxicity.

Le texte complet de cet article est disponible en PDF.

Mots-clés : Cytotoxicité, apoptose, cataracte, biocompatibilité, solution saline, sonde fluorescente, ISO 10993-5

Key-words : Cytotoxicity, apoptosis, cataract, biocompatible materials, saline solution, fluorescent probe, ISO 10993-5



© 2006  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 29 - N° 7

P. 773-780 - septembre 2006 Retour au numéro
Article précédent Article précédent
  • Traitement des néo-vaisseaux choroïdiens rétrofovéolaires du myope fort par photothérapie dynamique : résultats à 1 an et à 2 ans
  • L. El Matri, R. Bouraoui, A. Merdassi, R. Baccouri, F. Mghaieth, S. Ben Rejeb, S. Meddeb
| Article suivant Article suivant
  • Expression de la molécule d’adhésion intercellulaire de type 1 au niveau de la conjonctive des patients diabétiques : Une étude préliminaire
  • T. Khalfaoui, L. Kria, R. Anane, K. Erraies, O. Beltaief, G. Mkannez, R. Zhioua, S. Ben Jilani, A. Ouertani Meddeb

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2025 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.