J Fr. Ophtalmol., 2008; 31, 6, Cahier 2, 2S29-2S33
Intraocular pressure (IOP) can be measured through soft contact lenses with an air-puff tonometer. These measurements seemed accurate for low-power negative lenses. For positive soft contact lenses, IOP is overestimated. The measurement of IOP through a soft contact lens is acceptable in clinical practice only for glaucoma screening. In glaucomatous patients or patients with ocular hypertension, IOP measurements should be performed without a contact lens. The main technique for IOP measurement remains Goldmann applanation tonometry.
Le texte complet de cet article est disponible en PDF.Résumé |
La mesure de la pression intra-oculaire au travers des lentilles à renouvellement fréquent |
La pression intra-oculaire (PIO) peut être mesurée au travers des lentilles à renouvellement fréquent (LRF) à l’aide d’un tonomètre à air. Cette mesure semble fiable pour les lentilles de faibles puissances négatives. Pour les lentilles de puissances positives, cette mesure surestime la PIO. La mesure de la PIO au travers des lentilles de contact ne se conçoit en pratique clinique que dans le cadre du dépistage. Pour des patients glaucomateux ou avec une hypertonie oculaire, la mesure de la PIO doit se faire sans lentille de contact, la technique de référence étant la tonométrie de Goldmann.
Le texte complet de cet article est disponible en PDF.Keyword : Intraocular pressure, soft contact lenses, screening, air-puff tonometer
Mots-clés : Pression intra-oculaire, lentilles de contact souples, dépistage, tonomètre à air