Can we trust the accuracy of the automatic calibration of the EOS system to measure lower limb length inequality after total hip arthroplasty? Comparison of EOS versus manual measurement on 110 calibrated radiographs - 08/12/24

Doi : 10.1016/j.otsr.2024.104079 
Pierre Klein a, , Mehdi Hormi-Ménard b, Roger Erivan c, François Bonnomet a, Pablo Lamotte-Paulet a, Alain Duhamel d, Henri Migaud b

SOFCOTe

a Service de Chirurgie et de Traumatologie du Membre Inférieur, Hôpital Hautepierre 2, Hôpitaux Universitaires de Strasbourg, 67098 Strasbourg, France 
b Université de Lille Nord de France, Service d'orthopédie, Hôpital Roger-Salengro, Centre Hospitalier Régional Universitaire de Lille, Place de Verdun, 59037 Lille, France 
c Université Clermont Auvergne, CHU Clermont-Ferrand, CNRS, SIGMA Clermont, ICCF, F-63000 Clermont–Ferrand, France 
d University of Lille, CHU Lille, ULR2694 METRIX, Evaluation des Technologies de Santé et des Pratiques Médicales, 59000, Lille, France 
e Société Française de Chirurgie Orthopédique et de Traumatologique (SOFCOT), 56, rue Boissonade, 75014 Paris, France 

Corresponding author.
Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Sunday 08 December 2024

Abstract

Introduction

Traditionally, to determine a length on a limb radiograph after total hip arthroplasty (THA), calibration is performed manually with the diameter of a ball or the femoral head. More recently, the development of EOS with automatic calibration has called into question the usefulness of manual calibration to highlight lower limb length inequality (LLLI). However, the validation of EOS with automatic calibration without landmarks to measure length inequalities on large images has not been verified against manual measurements on calibrated radiographs (conventional method), which motivated the present work.

Hypothesis

EOS is more accurate and reproducible from one reader to another in highlighting LLLI after THA than the classic method.

Patients and methods

One hundred and ten patients included underwent primary THA surgery in 2 centers, with postoperative EOS performed. This EOS was extracted in 2 formats: a DICOM file with automatic calibration and an uncalibrated JPEG file (220 radiographic files in total).

Two readers, without knowledge of the clinical data, each analyzed all of these images using 2 methods: by measuring the distance between the center of the femoral head and the center of the mortise on the operated side and the non-operated side on the DICOM file, therefore via the software integrated into the EOS image (method 1), or by measuring these same distances on the JPEG file by calibrating the measurement to the diameter of the prosthetic head (method 2). The reproducibility of the measurements from one reader to another (inter-observer reproducibility) and the agreement of the measurements between the 2 methods (inter-method agreement) were evaluated using the intraclass correlation coefficient (ICC) and the Bland- Altman graphic method.

Results

Inter-method agreement was satisfactory regardless of the reader but with a significantly higher agreement for reader 1 (inter-method agreement for reader 1: CL1 = 0.95 and for reader 2: CL2 = 0.90 (p = 0.008)) and this result was also confirmed by the Bland-Altman plot with no bias tendency for each reader and less than 5% of measurements that were outside the agreement band. Inter-observer reproducibility for method 1 was better than that for method 2 to highlight an LLLI after THA according to the CCI. Inter-observer reproducibility for method 1 (RM1) was 0.96 and 0.92 for method 2 (RM2) (p = 0.009). This result was confirmed by the Bland-Altman plot with a mean difference of less than 1 mm for each of the methods (−0.2 (standard deviation = 2.23) and 0.81 (3.03) for method 1 and method 2 respectively).

Discussion

Our hypothesis has therefore been partially verified. The use of the calibrated software integrated into the EOS system and manual calibration are two valid methods for searching for an LLLI. The measurements remain more reproducible from one reader to another with the calibrated software integrated into the EOS system.

Level of evidence

III; non-randomized prospective comparative study.

Le texte complet de cet article est disponible en PDF.

Keywords : Leg length discrepancy, EOS imaging, X-rays, Hip prosthesis


Plan


© 2024  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2025 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.