Results of third generation dual mobility total hip arthroplasties with non-cross-linked polyethylene in patients under 60 years of age: comparative study of full pressfit vs. tripod cups after 8.6 years - 22/09/24

Doi : 10.1016/j.otsr.2024.103934 
Arthur Gras , Marianne Wiklund, Alexandre Ferreira, Valentin Chapus, Julien Dunet, Goulven Rochcongar, Christophe Hulet
 Département de Chirurgie Orthopédique du CHU de Caen Unité Inserm Comete 1075, Av. de la Côte de Nacre, 14000 Caen, France 

Corresponding author.

Abstract

Background

The concept of dual mobility of total hip arthroplasties (THA) is a revolution in the prevention of dislocations and in the treatment of instability. Its use remains controversial in patients under 60 years old, providing poorer results with first generation cups coated with alumina. This study, carried out on modern dual mobility (DM) cups with a porous bilayer coating, in active patients under 60 years of age aimed to evaluate: (1) the mechanical survival of two latest generation DM THA, with failure defined as revision of acetabular or femoral implants for mechanical loosening, by comparing a tripod cup to a full pressfit cup, (2) to evaluate the complications, (3) to compare the rate of radiographic peri-prosthetic osteolysis between the 2 implants.

Hypothesis

The hypothesis was that the medium-term survival of modern DM cups placed in young patients was the same regardless of the type of acetabular fixation used.

Materials and methods

This was a retrospective study of 119 latest generation DM THA with standard polyethylene, composed of 66 full pressfit cups and 53 tripod fixation cups, used in primary arthroplasty in 111 patients between 2005 and 2016. The survival study was carried out using medical records. The clinical evaluation was carried out using the modified Harris Hip Score (mHHS) as well as the self-administered Hip and Osteoarthritis Outcome Score (HOOS) and Oxford-12 questionnaires. Radiological analysis was carried out on radiographs looking for bone demineralization and acetabular and femoral lines.

Results

At 8.6 years (5–16), no revision for mechanical loosening was observed regardless of the type of cup. The prosthesis dislocation rate (large joint) was 1.7%: 1 case in each group (p = 1), including one revision required in the full pressfit group. Three revisions for mechanical complications were recorded: a recurrent dislocating hip on a full pressfit cup that was not sufficiently anteverted, an acetabular peri-prosthetic fracture on a full pressfit cup and a case of iliopsoas impingement with a retroverted tripod cup. The survival rate for all-cause revision was 93.8% for pressfit (95% CI: 88.2%–99.9%) vs. 96.2% for tripod (95% CI: 91.1%–100%) (p = 0.63) and for revision due to any cause excluding infection, 96.9% for pressfit (95% CI: 92.8%–100%) vs. 98.1% for tripod (95% CI: 94.4%–100%) (p = 0.7). The rate of periacetabular osteolysis was 11%, significantly more present amongst tripod cups (24%, 9 cases) compared to full pressfit cups (2%, 1 case) (p < 0.01).

Conclusion

This series demonstrates good performance of 3rd generation DM cups with porous bilayer coating in patients under 60 years of age regardless of the type of acetabular fixation, at 8.6 years of follow-up. The rate of osteolysis around full pressfit cups remains very low, unlike that of tripod cups. The use of Double Mobility prostheses for primary hip osteoarthritis is possible in young patients, preferably with full pressfit cups (i.e., without any pegs or additional screws that may promote diffusion of polyethylene debris and osteolysis).

Level of evidence

III; comparative retrospective study.

Le texte complet de cet article est disponible en PDF.

Keywords : Total hip arthroplasty, Dual-mobility cup, Survival, Young patients


Plan


© 2024  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 110 - N° 6

Article 103934- octobre 2024 Retour au numéro
Article précédent Article précédent
  • Change of joint line convergence angle and other coronal alignments after total hip arthroplasty
  • Ho Hyun Yun, Woo-seung Lee, Junwoo Park, Yun Seong Choi
| Article suivant Article suivant
  • The effect of body weight on interfragmentary fracture strain in plate fixation of distal femur fractures: A finite element analysis
  • Ishan D. Shah, Adam M. Schlauch, Lisa Phan, Jiho Han, Oluwatodimu Richard Raji M. Eng, Brian Farrell

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.