Evaluation of a Multimodal Confocal Therapeutic Focused Ultrasound Apparatus: Bridging Cavitation, Thermal Ablation, and Histotripsy in Preclinical Treatments - 10/09/24

Doi : 10.1016/j.irbm.2024.100861 
Myléva Dahan a, Maxime Lafond a, , R. Andrew Drainville a, Victor Delattre a, Marine Simonneau a, Françoise Chavrier a, Cyril Lafon a, Marion Cortet a, b, Frédéric Padilla a, c, d
a LabTAU, INSERM, Centre Léon Bérard, Université Claude Bernard Lyon 1, Univ Lyon, F-69003, LYON, France 
b Service de Gynécologie Obstétrique, Hôpital de la Croix Rousse, Hospices Civils de Lyon, F-69004, LYON, France 
c Focused Ultrasound Foundation, Charlottesville, VA, USA 
d Department of Radiology, University of Virginia School of Medicine, Charlottesville, VA, USA 

Corresponding author at: LabTAU, Inserm U1032, 151 Cours Albert Thomas, Lyon, 69003, France.LabTAUInserm U1032, 151 Cours Albert ThomasLyon69003France
Sous presse. Manuscrit accepté. Disponible en ligne depuis le Tuesday 10 September 2024

Abstract

Objectives

The development of versatile and user-friendly preclinical platforms is vital for therapeutic ultrasound research. We introduce a flexible ultrasound-guided focused ultrasound (FUS) platform with two confocal therapeutic transducers, allowing thermal and mechanical modalities, and present its design and, features, with validation of potential applications in preclinical studies.

Methods

The probe's acoustic properties, energy delivery efficiency, and thermal and mechanical modalities are characterized. A computational model predicts thermal effects while optimizing treatment parameters. Ex vivo tissue samples are used to validate system performance, safety, and usability. In vivo experiments on mice with MC38 tumors are presented with immunohistochemistry (IHC) to validate treatment outcomes.

Results

Electroacoustic conversion efficiency levels were 80% and 40% for 1.1 MHz and 3.3 MHz, respectively. Confocal therapy transducers at 1.1 MHz and 3.3 MHz successfully demonstrated cavitation histotripsy and thermal treatments. At 1.1 MHz, for histotripsy, −20 MPa negative peak pressure is achieved, while at 3.3 MHz used for thermal ablation a maximum of 35 MPa is reached for positive peak pressure. Numerical analysis provides thermal treatment planning, aligning with in vitro and in vivo experiments for lesion prediction. Real-time in vivo cavitation monitoring was consistent with in vitro chemical dosimetry, ensuring treatment uniformity.

Conclusion

The ultrasound platform induces thermal or mechanical lesions with precise spatial resolution, validated by IHC tissue characterization. Integrated cavitation monitoring enables real-time treatment monitoring. Coupling with thermal simulations provides optimization of thermal treatment parameters. This versatile”all-in-one” therapeutic platform supports multiple treatment modalities including cavitation, thermal ablation, and histotripsy, facilitating direct comparisons to assess their efficacy in diverse therapeutic settings.

Le texte complet de cet article est disponible en PDF.

Graphical abstract

Le texte complet de cet article est disponible en PDF.

Highlights

Versatile ultrasound platform using either the fundamental or the third harmonic.
Treatment modalities: Cavitation, thermal ablation, and histotripsy.
Simulations allowed prediction of thermal lesions.
Thermal and mechanical lesions were validated in vivo on histology.
Will allow comparison between ultrasound modalities, and their combination.

Le texte complet de cet article est disponible en PDF.

Keywords : Ultrasound platform, HIFU, cavitation, thermal ablation, histotripsy, preclinic device, confocal, multi modal therapeutic ultrasound


Plan


© 2024  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.