Analyse in silico de la zone de sécurité personnalisée d’antéversion de la cupule acétabulaire - 07/08/24

In silico analysis of the patient-specific acetabular cup anteversion safe zone

Doi : 10.1016/j.rcot.2024.07.011 
Thomas Aubert , Philippe Gerard, Giacomo Galanzino, Simon Marmor
 Croix Saint-Simon Hospital, Orthopaedic Department, 125, rue d’Avron, 75020 Paris, France 

Thomas Aubert, Croix Saint-Simon Hospital, Orthopaedic Department, 125, rue d’Avron, 75020 Paris, France.Croix Saint-Simon Hospital, Orthopaedic Department125, rue d’AvronParis75020France
Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Wednesday 07 August 2024

Résumé

Introduction

Différents systèmes de précision peropératoire prônent une amélioration de la précision de pose de la cupule dans les prothèses totales de hanches après analyse de la cinématique lombopelvienne afin d’éviter tout risque de conflit prothétique. Cependant aucune étude n’a analysé l’ampleur de la zone de sécurité d’antéversion de la cupule spécifique au patient.

Hypothèse

Notre hypothèse est que la majorité des patients ont une zone de sécurité de plus de 10° à l’exception des patients avec anomalie de la cinématique lombopelvienne où cette zone est beaucoup plus restreinte.

Matériel et méthodes

Nous avons simulé le risque de conflit prothétique en fonction de l’antéversion de la cupule planifiée. Notre cohorte consécutive comportait 341 patients opérés d’une prothèse totale de hanche. Notre critère de jugement principal était, pour chaque patient, la zone d’antéversion sans conflit puis divisée en quatre sous-groupes : 0°, 1–5°, 6–10° et>10°, secondairement analysée en fonction de la présence d’une anomalie de cinématique lombopelvienne (variation de tilt pelvien (ΔTP) de plus 20° de l’extension à la position assise).

Résultats

La zone de sécurité d’antéversion moyenne était de 22,8° avec un taux de 82,4 % (281/341) pour une zone strictement supérieure à 10°. Chez les patients avec un ΔTP20° (18,2 %), la zone de sécurité moyenne était de 8,9° (±9°), la proportion de patient avec une zone de 0°, 1–5°, 6–10° et>10° était respectivement de 37,1 %, 16,13 %, 8,06 % et 38,71 % et chez les patients avec un ΔTP<20° (81,8 %), la zone de sécurité moyenne était de 25,9° (±9°) et la proportion de patients était respectivement de 2,51 %, 1,08 %, 4,3 % et 92,11 % (p<0,001).

Conclusion

La zone de sécurité d’antéversion semble assez large chez une majorité de patients. En revanche l’identification des patients à risque d’anomalie de la cinématique lombopelvienne semble nécessaire pour dépister les deux tiers de patients avec une zone de sécurité restreinte.

Niveau de preuve

IV étude rétrospective.

Le texte complet de cet article est disponible en PDF.

Summary

Introduction

Various computer-assisted surgical systems claim to improve the accuracy of cup placement in total hip arthroplasties after assessing spinopelvic mobility to prevent prosthetic impingement. However, no study has yet analyzed the extent of the patient-specific cup anteversion safe zones.

Hypothesis

We hypothesized that most patients have a safe zone>10°, except those with abnormal spinopelvic mobility, who have a much narrower safe zone.

Materials and methods

We simulated the risks of prosthetic impingement using the planned cup anteversion. The consecutive cohort included 341 patients who underwent total hip arthroplasty. Our primary endpoint was the patient-specific impingement-free zone for cup anteversion, which was then divided into four subgroups: 0°, 1° to 5°, 6° to 10°, and>10°. This data was then secondarily analyzed for abnormal spinopelvic mobility (the difference in the spinopelvic tilt [ΔSPT] from a standing to a flexed seated position>20°).

Results

The mean anteversion safe zone was 22.8° with 82.4 % (281/341) of patients with a zone strictly>10°. The mean safe zone was 8.9° (±9°) in patients with an ΔSPT20° (18.2 %), with 37.1 % of these patients having a zone of 0°, 16.13 % a zone between 1° and 5°, 8.06 % a zone between 6° and 10° and 38.71 % a zone>10°. The mean safe zone was 25.9° (±9°) in patients with an ΔSPT<20° (81.8 %), and the proportion of cases in each zone was 2.51 %, 1.08 %, 4.3 %, and 92.11 %, respectively (p<0.001).

Conclusion

The safe zone for anteversion appears to be fairly wide in most patients. However, identifying patients at risk of abnormal spinopelvic mobility seems necessary to identify the two-thirds of patients with a narrow safe zone.

Level of evidence

IV; retrospective study.

Le texte complet de cet article est disponible en PDF.

Mots clés : PTH, Conflit prothétique, Cinématique lombopelvienne, Relation hanche rachis

Keywords : Total hip arthroplasty, THA, Prosthetic impingement, Spinopelvic mobility, Hip-spine relationship


Plan


 Ne pas utiliser, pour citation, la référence française de cet article, mais celle de l’article original paru dans Orthopaedics & Traumatology: Surgery & Research, en utilisant le DOI ci-dessus.


© 2024  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.