Abstract

Introduction

In revision total hip arthroplasty (THA), the advent of porous custom-made triflange acetabular implants with 3D scan planning offers a new perspective to improve implantation accuracy and anatomical restoration of the center of rotation (COR). This issue was investigated using CT-scan as the measurement tool, but in limited series (±10 cases) and without investigating the factors that may influence errors in positioning. Therefore we performed a retrospective study aiming to: (1) assess the placement accuracy of such implants with respect to the preoperative planning, (2) examine whether the volume of bone to be resected in order to apply the implant had an impact on this accuracy, (3) assess if errors in position at surgery had any influence on function, complications and survival.

Hypothesis

Preoperative planning could be accurately reproduced when implanting porous custom-made acetabular implants, and that accuracy would decrease in proportion to the volume of bone to be resected

Method

Twenty patients undergoing THA revision with porous custom-made acetabular implants were included in this single-center retrospective study. Mean follow-up was 17.9 months ± 9.4 [2–45.1]. Preoperative planning was performed using 3D scanographic modeling. A post-operative CT scan was performed to assess implantation accuracy in terms of orientation and COR restitution. Demographic data, Oxford scores, complications and survival were recorded.

Results

Mean deviation from the preoperative planning in inclination, anteversion and rotation were 4.3 ° ± 2.5, 6.1 ° ± 4.7, and 7 ° ± 4.6, respectively. Restoration of the COR showed a mean deviation of 2.1 ± 1.3 mm anteroposteriorly, 2.5 ± 2 mm mediolaterally and 2.2 ± 1.3 mm proximodistally. In total, 45% (9/20) of implants were positioned with perfect restoration of orientation (±10 °) and COR (±5 mm). The mean planned bone resection was 8.1 ± 4.9 cm3, with placement accuracy and COR restitution decreasing significantly when the volume of bone to be resected exceeded 2.7 cm3. One dislocation was found (5%, 1/20). Survival at last follow-up was 100%, the mean Oxford score at follow-up was 31.7 ± 7.9 [16–52], without being influenced by errors in position or COR restitution.

Conclusion

In total 45% of the implants restored an orientation and a COR as planned, particularly when the volume of bone to be resected is less than 2.7 cm3. Although these are complex cases with large amounts of bone loss, 3D manufacturing could give us hope of greater precision. The link between better precision and low bone resection volume could be an area to develop with the manufacturer in order to improve results.

Level of evidence

III; diagnostic using CT in transversal retrospective study

Le texte complet de cet article est disponible en PDF.

Keywords : Acetabular revision, Accuracy, Large defect, Custom-made, 3D printed, 3D planning


Plan


© 2024  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.