S'abonner

How to be sure of a complete removal of Essure® implant? A radiological study of Essure® microinserts - 25/07/24

Doi : 10.1016/j.jogoh.2024.102823 
Gautier Chene a, b, , Charlotte Formont a, Aphynia Sacdpraseuth a, Lea Puyol a, Chloe Miguet-Bensouda a, Stephanie Moret a, Emanuele Cerruto a, Erdogan Nohuz a
a Department of Gynecology, Hôpital Femme Mère Enfant, HFME, University Hospital of Lyon, 59 boulevard Pinel, Lyon 69000, France 
b Claude Bernard University of Lyon 1, EMR 3738, Lyon 69000, France 

Corresponding author at: Department of Gynecology, Hôpital Femme Mère Enfant, HFME, University Hospital of Lyon, 59 boulevard Pinel, Bron 69500, France.Department of Gynecology, Hôpital Femme Mère Enfant, HFMEUniversity Hospital of Lyon59 boulevard PinelBron69500France

Abstract

Purpose

Laparoscopic cornuectomy, one of the most performed surgical techniques to remove Essure® implants, is associated with a risk of fragmentation of the device. We aimed to assess the diagnostic accuracy of the intraoperative radiograph of the removed Essure® microinserts compared with intraoperative inspection of the device.

Study design

In this French retrospective cohort study including 66 patients, two experienced gynecologists and two residents evaluated the intraoperative radiograph in a blind manner. Diagnostic accuracy was calculated for each observer.

Results

The global sensitivity and specificity for intraoperative radiograph to detect a residual fragment for both senior gynecologists and residents were, respectively 87.5 % (±3.3) and 80.0 % (+/12.2). Pooled positive (PPV) and negative predictive values (NPV) for intraoperative radiograph were respectively 99.2 % (±0.5) and 21.6 % (±6.4). There was a statistically significant difference in intraoperative radiograph analysis between senior gynecologists and residents (sensitivity: 91.4 %±0.8 vs 79.6 %±1.5, p = 0.02; specificity: 100% vs 50.0 %, p < 0.0001; PPV: 100 % vs 98.0 % ±0.1, p = 0.0007; NPV: 26.8 % ±1.8 vs 7.2 % ±0.5, p = 0.009).

Sensitivity and specificity of the intraoperative inspection of the removal specimen was respectively 95.3 % and 100 % (PPV and NPV were, respectively 100 and 40 %).

There were no statistically significant differences in diagnostic accuracy between intraoperative radiographs analysis and intraoperative inspection of the device for the senior gynecologists.

Conclusions

Intraoperative radiograph of the removed Essure® implants analysis is a difficult skill needing specific training. Postoperative pelvic X-ray is still recommended in clinical practice after laparoscopic cornuectomy.

Le texte complet de cet article est disponible en PDF.

Keywords : Essure® removal, X-ray, Fallopian tube, Cornuectomy, Pelvic radiograph


Plan


 Short condensation: X-ray interpretation of the removed Essure® implants is a difficult skill needing specific training.


© 2024  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 53 - N° 9

Article 102823- novembre 2024 Retour au numéro
Article précédent Article précédent
  • Intriguing connection between diminished ovarian reserve and childhood traumatic experiences, a prospective case-control study
  • Reyhan Çanakci, Yeşim Uncu, Kiper Aslan, Işıl Kasapoğlu, Gürkan Uncu
| Article suivant Article suivant
  • Case study: Recurrent uterine incarceration during pregnancy
  • Laetitia Roussel, Callyane Seve, Romain Cahierc, Fanny Petillon, Martine Bonnin, Amélie Delabaere, Marion Rouzaire, Denis Gallot

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.