S'abonner

Analyse d’imputabilité d’un ciment basse viscosité en orthopédie suite à la survenue d’événements indésirables graves - 19/10/23

Causality analysis of a low-viscosity bone cement in orthopaedic surgery following serious adverse events

Doi : 10.1016/j.pharma.2023.03.006 
Amélie Cherpin a, , Marion Poirier b, Jean-Camille Mattei c, Bernard Lassale d, Florence Peyron a
a Service pharmacie, hôpital Nord Marseille (AP–HM), chemin des Bourelly, 13015 Marseille, France 
b Département d’anesthésie-réanimation, hôpital Nord Marseille (AP–HM), Marseille, France 
c Département de chirurgie orthopédique et traumatologie, hôpital Nord Marseille (AP–HM), Marseille, France 
d Direction de la qualité et de la gestion des risques, AP–HM, Marseille, France 

Auteur correspondant.

Résumé

Points essentiels

Prise en charge chirurgicale de patients avec comorbidités.
Bénéfices des revues de morbi-mortalité et des analyses de matériovigilance.
Risques de l’utilisation du ciment osseux en chirurgie orthopédique.
Assurance qualité et mise en place d’actions correctives
Intérêts de l’analyse pluridisciplinaire.

Le texte complet de cet article est disponible en PDF.

Résumé

Objectifs

Le ciment osseux basse viscosité imprégné de gentamicine est fréquemment utilisé pour la fixation des prothèses fémorales. Trois arrêts cardiaques sur cimentoplastie ont eu lieu successivement sur des interventions de prothèse de hanche, entraînant le décès de deux patientes. L’objectif de l’analyse est d’exposer les actions engagées afin d’établir un lien potentiel entre l’utilisation du ciment et la survenue de ces événements indésirables graves (EIG).

Méthodes

Une revue de mortalité et de morbidité a été organisée afin d’étudier l’imputabilité du ciment et de proposer des actions d’amélioration suite à trois EIG déclarés et associés à des déclarations de matériovigilance.

Résultats

Les trois EIG sont survenus à la suite de l’injection de la même référence de ciment. Les lots incriminés ont été mis en quarantaine sans délai. L’analyse du fabricant n’a pas relevé de manquement aux exigences qualités de production mais a suggéré la survenue d’un syndrome d’implantation au ciment osseux (BCIS). Une revue de la littérature sur le BCIS confirme que cette complication peropératoire rare était plausible pour les trois cas. La gestion de ces EIG inscrite dans une démarche sécurité des soins a permis de statuer rapidement sur l’imputabilité du ciment et les écarts de pratique associés à son utilisation.

Conclusions

L’analyse systémique complétée par l’analyse du fabricant a dégagé des mesures correctives concernant les pratiques professionnelles. La réalisation et l’efficacité des actions engendrées feront l’objet d’un suivi dans le programme d’amélioration de la qualité et de la sécurité des soins de l’établissement.

Le texte complet de cet article est disponible en PDF.

Summary

Objectives

Low-viscosity bone cement impregnated with gentamicin is frequently used to fix femoral prostheses. Three cardiac arrests occured successively during cementoplasty oh hip replacements, leading to the death of two patients. The objective of this study is to describe the actions undertaken to establish a potential link between the use of the bone cement and the occurrence of these serious adverse events (SAE).

Methods

A mortality and morbidity review was organised in order to study the causality of bone cement and to propose improvement actions, following 3 considered SAE associated to materiovigilance reporting.

Results

All three SAE occurred following the injection of the same reference of bone cement. The incriminated batches were rapidly placed in quarantine. Analysis by the manufacturer revealed no defects in production quality requirements but suggested the possibility of Bone Cement Implantation Syndrome (BCIS). A literary review on BCIS confirmed that this rare intraoperative complication was plausible in all three cases. Management of these SAE via a health care safety process enabled to provide a rapid answer concerning the causality of the cement and practice deviations of its use.

Conclusions

Systemic analysis completed by the manufacturer's analysis provided corrective actions for professional practices. Implementation and efficacy of these actions will be monitored as part of the facility's programme for the improvement of quality and patient safety.

Le texte complet de cet article est disponible en PDF.

Mots clés : Ciments osseux, Orthopédie, Evénement indésirable grave associé aux soins

Keywords : Bone cement, Orthopaedics, Serious adverse incident associated with health care


Plan


© 2023  Académie Nationale de Pharmacie. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 81 - N° 6

P. 1031-1037 - novembre 2023 Retour au numéro
Article précédent Article précédent
  • Premières journées francophones des utilisateurs de système d’aide à la décision en pharmacie clinique : retour d’expériences et perspectives
  • L. Robert, C. Rousselière, J.-B. Beuscart, S. Gautier, L. Delporte, G. Lafci, E. Gerard, L. Négrier, A. Mary, E. Johns, A. Payen, R. Ducommun, L. Ferret, P. Voirol, C. Skalafouris, M. Ade, A. Potier, E. Dufay, J. Beney, Pauline Frery, Sylvain Drouot, F. Feutry, J. Corny, P. Odou, B. Décaudin
| Article suivant Article suivant
  • Circuit pharmaceutique du médicament de thérapie innovante (MTI) en France : état des lieux et perspectives
  • Benjamin Fabri, Guillaume Sicard, Laurence Gauthier-Villano, Bertrand Pourroy

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2025 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.