Agreement of Oscillometric and Auscultatory blood pressure measurement methods: An ambulance noise simulation study - 20/04/23

Doi : 10.1016/j.ajem.2023.02.022 
Ali Cankut Tatliparmak a, Sarper Yilmaz b,
a Dept. of Emergency Medicine, Kolan International Hospital, Istanbul, Turkey 
b University of Health Sciences, Dept. of Emergency Medicine, Kartal Dr. Lutfi Kirdar City Hospital, Istanbul, Turkey 

Corresponding author at: Department of Emergency Medicine, University of Health Sciences, Kartal Dr. Lutfi Kirdar City Hospital, Kartal, Istanbul, Turkey.Department of Emergency MedicineUniversity of Health SciencesKartal Dr. Lutfi Kirdar City HospitalIstanbulKartalTurkey

Abstract

Objective

Although noise is known to negatively affect blood pressure (BP) measurements, its impact on different BP measurement methods remains unclear. The aim of this study is to compare the agreement of oscillometric and auscultatory BP measurement methods under in-ambulance noise levels.

Methods

This method-comparison study was conducted on 50 healthy volunteers in a tertiary emergency department (ED). Participants were divided into two groups of 25, and BP was measured using auscultatory and oscillometric methods in noisy and ambient environments by 2 emergency medicine technicians (EMT). The primary object of the study was to compare the agreement of auscultatory mercury sphygmomanometers and automated auscillometric BP measurements in ambient and noisy environments.

Results

We examined the agreement between auscultative and oscillometric measurements of BP conducted in an ambient environment (46.75 [IQR (41.2--55.18)] dB) and found that both systolic and diastolic BP were within the level of agreement (LoA) established before the study (systolic BP [−13.96 to 8.48 mmHG], diastolic BP [−7.44 to 8.08 mmHg]); whereas, in noisy environment (92.35 [IQR 88–96.55] dB) both systolic and diastolic BP were outside the range of LoA (systolic BP [−37.77 to 9.94 mmHg], diastolic BP [−21.73 to 16.37 mmHg]). Additionally, we found that in ambient environments, concordance correlation coefficients were higher than in noisy environments (0.943 [0.906–0.966], 0.957 [0.93–0.974]; 0.574 [0.419–0.697], 0.544 [0.326–0.707]; systolic and diastolic BP, respectively).

Conclusion

The results of this study demonstrate that noise significantly affects the agreement between oscillometric and auscultatory blood pressure measurement methods.

Le texte complet de cet article est disponible en PDF.

Keywords : Blood pressure measurement method, Oscillometric blood pressure measurement, Auscultatory blood pressure measurement, Ambulance noise


Plan


© 2023  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 67

P. 120-125 - mai 2023 Retour au numéro
Article précédent Article précédent
  • Bladder deformity accompanied by pelvic fracture indirectly indicates clinical severity
  • Soichiro Ota, Ikuto Takeuchi, Michika Hamada, Wataru Fujita, Ken-ichi Muramatsu, Hiroki Nagasawa, Kei Jitsuiki, Hiromichi Ohsaka, Kouhei Ishikawa, Atsuhiko Mogami, Youichi Yanagawa
| Article suivant Article suivant
  • Acute rate control with metoprolol versus diltiazem in atrial fibrillation with heart failure with reduced ejection fraction
  • Karolina Kapustova, Brian Phan, Timothy Allison-Aipa, Mia Choi

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.