Aerosol containment device design considerations and performance evaluation metrics - 13/01/23

Doi : 10.1016/j.ajem.2022.11.007 
Rachael M. Jones, PhD a, b, , Niles Andrus, MS a, Thomas Dominguez, MS a, Jeremy Biggs, MD a, Brian Hansen, MS a, Frank A. Drews, PhD c
a Department of Family and Preventive Medicine, Spencer Fox Eccles School of Medicine, University of Utah, United States of America 
b Department of Environmental Health Sciences, Fielding School of Public Health, University of California, Los Angeles, United States of America 
c Department of Psychology, College of Social and Behavioral Science, University of Utah, United States of America 

Corresponding author at: 650 Charles E Young Dr. S, 71-295, Center for Health Sciences, Los Angeles, CA 90095, United States of America.Center for Health Sciences650 Charles E Young Dr. S, 71-295Los AngelesCA90095United States of America

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Background

Spurred by the Coronavirus infectious disease 2019 pandemic, aerosol containment devices (ACDs) were developed to capture infectious respiratory aerosols generated by patients at their source. Prior reviews indicated that such devices had low evidence of effectiveness, but did not address how ACDs should be evaluated, how well they should perform, nor have clearly defined performance standards. Towards developing design criteria for ACDs, two questions were posed: 1) What characteristics have guided the design of ACDs? 2) How have these characteristics been evaluated?

Methods

A scoping review was performed consistent with PRISMA guidelines. Data were extracted with respect to general study information, intended use of the device, device design characteristics and evaluation.

Results

Fifty-four articles were included. Evaluation was most commonly performed with respect to device aerosol containment (n = 31, 61%), with only 5 (9%), 3 (6%) and 8 (15%) formally assessing providing experience, patient experience and procedure impact, respectively. Nearly all of the studies that explored provider experience and procedure impact studied intubation. Few studies provided a priori performance criteria for any evaluation metric, or referenced any external guidelines by which to bench mark performance.

Conclusion

With respect to aerosol containment, ACDs should reduce exposure among HCP with the device compared with the absence of the device, and provide ≥90% reduction in respirable aerosols, equivalent in performance to N95 filtering facepiece respirators, if the goal is to reduce reliance on personal protective equipment. The ACD should not increase awkward or uncomfortable postures, or adversely impact biomechanics of the procedure itself as this could have implications for procedure outcomes. A variety of standardized instruments exist to assess the experience of patients and healthcare personnel. Integration of ACDs into routine clinical practice requires rigorous studies of aerosol containment and the user experience.

Le texte complet de cet article est disponible en PDF.

Keywords : Aerosol box, COVID−19, Engineering controls, Respiratory aerosols

Abbreviations : ACD, AIIR, APF, COVID-19, D, FD, FFR, HCP, M, NASA TLX, PRISMA, R, SARS-CoV-2


Plan


© 2022  The Authors. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 64

P. 12-20 - février 2023 Retour au numéro
Article précédent Article précédent
  • Recent developments and controversies in therapeutic hypothermia after cardiopulmonary resuscitation
  • Peijuan Li, Zhangping Sun, Tian Tian, Dongping Yu, Hui Tian, Ping Gong
| Article suivant Article suivant
  • Continuous compression with asynchronous ventilation improves CPR prognosis? A meta-analysis from human and animal studies
  • Mengxue Sun, Aiqun Zhu, Yangyang Tang

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.