S'abonner

Pandemic driven innovation: A pilot evaluation of an alternative respiratory pathogen collection device - 18/10/22

Doi : 10.1016/j.ajem.2022.08.047 
Thang T. Nguyen, PhD a , Wesley G. Zeger, DO a , Michael C. Wadman, MD a , Andy T. Schnaubelt, PhD b , Aaron N. Barksdale, MD a,
a Department of Emergency Medicine, United states 
b Department of Neurosurgery, United states 

Corresponding author at: The University of Nebraska Medical Center, 981150, Nebraska Medical Center, Omaha, NE 68198-1150, United States of America.The University of Nebraska Medical CenterNebraska Medical Center981150OmahaNE68198-1150United States of America

Abstract

Background

The nasopharyngeal swab is the gold standard collection method for COVID-19, but is invasive and painful, subsequently resulting in poor patient acceptance. This investigation explores the process of developing and validating an alternative respiratory pathogen collection device that relies on a nasopharyngeal irrigation mechanic. The primary objective was to determine if sufficient pathological sampling can be achieved by mechanism of nasopharyngeal irrigation that is proportionate to the nasopharyngeal swab method.

Methods

The study device was designed using Shapr3D modeling software and fabricated on a fused deposition modeling printer. Fifteen participants were enrolled with each receiving a saline nasopharyngeal washing using the study device. Specimen adequacy was evaluated by two real-time reverse transcriptase polymerase chain reaction (PCR) testing methods to identify the presence of the human RNase P gene. Results were evaluated quantitatively through interpretation of the PCR cycle threshold (Ct).

Results

All 15 specimens tested positive for the presence of RNaseP, demonstrating specimen cellularity, adequate extraction of nucleic acids, and the absence of inhibitors to amplification. The mean Ct value was 29.5 (Applied Biosystems TaqPath RT-qPCR) and 30.7 (NECoV19). All participants felt the study device irrigation procedure was faster than the nasopharyngeal swab, with none experiencing any discomfort from the irrigation mechanism.

Conclusion

The importance of early diagnostic testing and its role in countermeasures for communicable diseases such as COVID-19 is well established in the literature. Innovation to bolster our testing infrastructure is more important now than ever. This study was successful in developing and validating an alternative nasopharyngeal respiratory pathogen collection device that utilizes fluid debridement as its core mechanic. Data from this pilot study demonstrated the study device was successful in producing high-quality specimens for PCR testing. Feedback from the study participants was also in favor of the study device when compared to the nasopharyngeal swab.

Le texte complet de cet article est disponible en PDF.

Keywords : Respiratory virus, Self-collection respiratory pathogen device, COVID-19


Plan


 Moderated Poster presentation: Society of Academic Emergency Medicine (SAEM) Annual Meeting, New Orleans, Louisiana, May 13, 2022.


© 2022  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 61

P. 111-116 - novembre 2022 Retour au numéro
Article précédent Article précédent
  • Covered or uncovered: A randomized control trial of Tegaderm versus no Tegaderm for ocular ultrasound
  • Amy Marks, Daven Patel, Mark Chottiner, Fae Kayarian, Gary D. Peksa, Michael Gottlieb
| Article suivant Article suivant
  • Post-traumatic symptoms in patients with acute coronary syndrome: Maybe an outcome predictor after the ED visit!
  • Noureddine Rekik, Sirine Bouzid, Amine Abdelhedi, Kais Bouzid, Mouna Benamor, Feres Benamira, Rim Karray, Mouna Jerbi, Abdennour Nasri, Olfa Chakroun-Walha

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.